Knowledge base

The Lab

Try measurement technology right in the browser. The simulations are based on tabulated nuclide data and the manufacturers’ own specifications – and say where they simplify.

Spectrum lab

Same radiation – two detectors

Choose one or more nuclides and two detectors. The spectrum builds up pulse by pulse, and both detectors receive the same photons – the difference you see is the resolution.

Live time 00:00:00Pulses 0

A · NaI(Tl)≤ 7.2% FWHM at 662 keVPeak width at 121.8 keV ≈ 20.4 keV
B · HPGe≤ 0.87 keV at 122 keV; ≤ 1.8 keV at 1.33 MeVPeak width at 121.8 keV ≈ 0.87 keV
NaI(Tl)

Rugged, cost-effective scintillator – common in handheld identifiers and fixed monitors.

HPGe

Highest resolution – requires cooling with liquid nitrogen or an electro-mechanical cooler.

The lines in the simulation and their sources
NuclideEnergy (keV)Photons per 100 decaysType
Eu-15239.522920.8K X-ray
Eu-15240.118637.7K X-ray
Eu-15242.30930.243K X-ray
Eu-15242.99670.437K X-ray
Eu-15245.41311.78K X-ray
Eu-15246.5753.04K X-ray
Eu-15248.6970.138K X-ray
Eu-152121.781728.41gamma
Eu-152244.69747.55gamma
Eu-152295.93870.442gamma
Eu-152329.4250.129gamma
Eu-152344.278526.59gamma
Eu-152367.78910.862gamma
Eu-152411.11652.238gamma
Eu-152416.0480.109gamma
Eu-152443.9653.12gamma
Eu-152488.67920.4139gamma
Eu-152503.4740.1533gamma
Eu-152563.990.457gamma
Eu-152566.4420.131gamma
Eu-152586.2650.462gamma
Eu-152656.4890.1437gamma
Eu-152674.6750.17gamma
Eu-152678.6230.47gamma
Eu-152688.670.841gamma
Eu-152719.3490.268gamma
Eu-152764.90.19gamma
Eu-152778.904512.97gamma
Eu-152810.4510.317gamma
Eu-152841.5740.163gamma
Eu-152867.384.243gamma
Eu-152919.3370.429gamma
Eu-152926.3170.273gamma
Eu-152963.390.1341gamma
Eu-152964.07914.5gamma
Eu-1521,005.2720.665gamma
Eu-1521,085.83710.13gamma
Eu-1521,089.7371.73gamma
Eu-1521,109.1740.186gamma
Eu-1521,112.07613.41gamma
Eu-1521,212.9481.416gamma
Eu-1521,249.9380.186gamma
Eu-1521,292.7780.104gamma
Eu-1521,299.1421.633gamma
Eu-1521,408.01320.85gamma
Eu-1521,457.6430.498gamma
Eu-1521,528.1030.281gamma

Simulation for illustration. Line energies and intensities are from LNHB/CEA, Table de Radionucléides: gamma lines with at least 0.1 photons per 100 decays and K X-rays above 24 keV. Peak widths are derived from the manufacturer’s resolution figure – scaled with √E from 662 keV for scintillators and CZT, fitted to the values at 122 keV and 1.33 MeV for HPGe. Efficiency and Compton background are simplified and identical for both detectors, and all selected nuclides have the same activity. In a real instrument the crystal size, the window and the measurement geometry also shape the spectrum.