Guide

How to read a data sheet – where a value comes from and what it promises

A value in a data sheet applies under stated conditions. Here is how to find the source of a figure, tell guaranteed limits from typical values and compare instruments on equal terms.

Step by step

One value, from catalogue to comparison

Scroll to follow one of the GEM40’s values.

Relative efficiency40 % (minimum)1
Energy resolution at 122 keV≤ 0.87 keV FWHM1
Energy resolution at 1.33 MeV≤ 1.8 keV FWHM1
Peak-to-Compton ratio≥ 64:11
Peak shape FW.02M/FWHM2.6 (typical)1
  1. 01

    The value

    Every technical value in the catalogue has a numbered source. The example is the ORTEC GEM40, an HPGe detector with at least 40 % relative efficiency.

  2. 02

    Where it is written

    The number leads to the manufacturer’s document – here p. 18 of the GEM series configuration guide, the table “Ordering Information and Warranted Specifications”. You can check the value there yourself.

  3. 03

    What it promises

    In the table, ≤, ≥ and minimum are guaranteed limits. A typical value – such as the peak shape of 2.6 – is the normal outcome, not a guarantee. Resolution also applies at a stated energy, here 122 keV and 1.33 MeV.

  4. 04

    Compare on equal terms

    The GMX40 also has at least 40 %, but at most 2.0 keV at 1.33 MeV against the GEM40’s 1.8 keV – and the GMX series measures from about 5 keV instead of about 40 keV. The comparison shows the differences straight away and can be downloaded as PDF or Excel, with the sources.

Where the value comes from

In NDEC’s catalogue every technical value has a numbered source. The number after the value refers to the source list further down the product page, where the manufacturer’s document is linked and the location of the value in it is given. For the GEM40, the relative efficiency of 40 % (minimum) is on p. 18 of ORTEC’s configuration guide for the GEM series – in the table Ordering Information and Warranted Specifications.

Guaranteed or typical

Manufacturers state values in different ways. A limit – a maximum (≤) or minimum (≥) – is a guarantee. A typical value is the normal outcome, but not a guarantee.

In ORTEC’s table for the GEM series, resolution and peak-to-Compton ratio are guaranteed limits: the GEM40 has ≤ 0.87 keV FWHM at 122 keV, ≤ 1.8 keV FWHM at 1.33 MeV and a peak-to-Compton ratio of ≥ 64:1. The same table gives the peak shape FW.02M/FWHM as a typical value: 2.6 (typical). In the catalogue the value keeps the manufacturer’s wording – minimum, maximum or typical.

The conditions a value applies under

A value applies at a stated energy, temperature or measurement geometry. For HPGe, resolution is given at 122 keV and 1,332 keV; for scintillators and CZT usually at 662 keV – so only compare values that apply at the same energy. Relative efficiency is a comparison number defined in IEEE Std 325 for a fixed geometry; read more in the guide Relative efficiency of HPGe detectors.

Compare on equal terms

The GMX40 has the same relative efficiency as the GEM40, 40 % (minimum). Still they differ: the GMX40 has ≤ 2.0 keV FWHM at 1.33 MeV against the GEM40’s ≤ 1.8 keV FWHM, and the GMX series measures from about 5 keV while the GEM series is used from about 40 keV and up. Differences like these only show when the values sit side by side under the same conditions.

In the comparison you can put up to four products side by side, show only the differences and download the table as PDF or Excel – with the sources.

Checklist

  • Source: which document does the value come from, and does it apply to the model you are looking at?
  • Limit or typical: does it say ≤, ≥, minimum, maximum or typical?
  • Conditions: at which energy, temperature or geometry does the value apply?
  • Up to date: check against the manufacturer’s latest data sheet before ordering – the exact configuration is always confirmed in the quotation.

References

  1. ORTEC: GEM Series Coaxial HPGe Detector – Product Configuration Guide, p. 18
  2. ORTEC: GEM P-type Coaxial HPGe Radiation Detectors (product page)
  3. ORTEC: GMX Series Coaxial HPGe Detector – Product Configuration Guide, p. 14
  4. ORTEC: Gamma-X (GMX) N-type HPGe Coaxial Radiation Detectors (product page)
  5. IEEE Std 325-1996: IEEE Standard Test Procedures for Germanium Gamma-Ray Detectors

Unfamiliar term? Look it up in the glossary.

Product areas

Relevant products

GEM HPGe-detektorer
ORTEC

GEM HPGe-detektorer

HPGe detectors for gamma spectroscopy

Energy range
Approximately 40 keV and above
Relative efficiency
10–150% across the model series; higher on request
Energy resolution at 122 keV
FWHM maximum limits 0.80–1.30 keV for GEM10–GEM150; model-dependent
17 models
GAMMA-X (GMX) HPGe-detektorer
ORTEC

GAMMA-X (GMX) HPGe-detektorer

HPGe gamma detector

Energy range
Approximately 5 keV–10 MeV
Relative efficiency
10–100%; higher on request
Energy resolution at 122 keV
FWHM maximum limits 0.85–1.40 keV for GMX10–GMX100; model-dependent; optimal shaping with DSPEC-50A
12 models

More guides

Knowledge base
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.