GEM HPGe-detektorer

HPGe detectors for gamma spectroscopy

P-type coaxial germanium detectors for laboratory measurements from approximately 40 keV and above. Select a configuration for the sample geometry and measurement task.

Energy range
Approximately 40 keV and above1
Relative efficiency
10–150% across the model series; higher on request2
Energy resolution at 122 keV
FWHM maximum limits 0.80–1.30 keV for GEM10–GEM150; model-dependent2
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Relative efficiency up to 150%, higher on request.
  • PopTop, Streamline and mechanical cooling configurations.
  • SMART-1 bias and carbon fibre endcaps are available as options.

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Detector typeP-type coaxial HPGe1

Performance

Relative efficiency10–150% across the model series; higher on request2
Energy resolution at 122 keVFWHM maximum limits 0.80–1.30 keV for GEM10–GEM150; model-dependent2
Energy resolution at 1.33 MeVFWHM maximum limits 1.7–2.3 keV for GEM10–GEM150; model-dependent2
Resolution measurement conditionsTotal system resolution measured at the factory per ANSI/IEEE Std 325-1996, at optimal shaping time with an ORTEC DSPEC-50A2

Measuring range

Energy rangeApproximately 40 keV and above1

Electronics & signal

PreamplifierOptional PLUS for high count rates1

Cooling

CoolingLN₂ or compatible mechanical cooling, by configuration1

Mechanical

EndcapLow-background carbon fibre option1

Ordering & accessories

Cryostat configurationsPopTop or Streamline1
Bias controlSMART-1 (-SMP), optional1
Cooling and accessoriesCryostat and dewar or other cooling device required for operation and not included with the detector; a cryostat must be ordered with a Streamline detector2

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Models

17 models in the series. Filter or scroll the table to compare.

17 / 17
ModelRelative efficiency2Energy resolution at 122 keV2Energy resolution at 1.33 MeV2Peak-to-Compton ratio2Peak shape FW.1M/FWHM2Peak shape FW.02M/FWHM2Endcap diameter2Ordering configurations2
GEM1010 % (minimum)≤ 0.80 keV FWHM≤ 1.7 keV FWHM≥ 41:11.9 (maximum)2.6 (typical)70 mm (nominal)GEM10P4-70 (PopTop) / GEM10-70 (Streamline)
GEM1515 % (minimum)≤ 0.82 keV FWHM≤ 1.8 keV FWHM≥ 46:11.9 (maximum)2.6 (typical)70 mm (nominal)GEM15P4-70 (PopTop) / GEM15-70 (Streamline)
GEM2020 % (minimum)≤ 0.82 keV FWHM≤ 1.8 keV FWHM≥ 52:11.9 (maximum)2.6 (typical)70 mm (nominal)GEM20P4-70 (PopTop) / GEM20-70 (Streamline)
GEM2525 % (minimum)≤ 0.82 keV FWHM≤ 1.8 keV FWHM≥ 56:11.9 (maximum)2.6 (typical)76 mm (nominal)GEM25P4-76 (PopTop) / GEM25-76 (Streamline)
GEM3030 % (minimum)≤ 0.85 keV FWHM≤ 1.8 keV FWHM≥ 60:11.9 (maximum)2.6 (typical)76 mm (nominal)GEM30P4-76 (PopTop) / GEM30-76 (Streamline)
GEM4040 % (minimum)≤ 0.87 keV FWHM≤ 1.8 keV FWHM≥ 64:11.9 (maximum)2.6 (typical)76 mm (nominal)GEM40P4-76 (PopTop) / GEM40-76 (Streamline)
GEM5050 % (minimum)≤ 0.90 keV FWHM≤ 1.9 keV FWHM≥ 66:11.9 (maximum)2.6 (typical)83 mm (nominal)GEM50P4-83 (PopTop) / GEM50-83 (Streamline)
GEM6060 % (minimum)≤ 1.00 keV FWHM≤ 1.9 keV FWHM≥ 70:11.9 (maximum)3.0 (typical)95 mm (nominal)GEM60P4-95 (PopTop) / GEM60-95 (Streamline)
GEM7070 % (minimum)≤ 1.00 keV FWHM≤ 2.0 keV FWHM≥ 75:12.0 (maximum)3.0 (typical)95 mm (nominal)GEM70P4-95 (PopTop) / GEM70-95 (Streamline)
GEM8080 % (minimum)≤ 1.10 keV FWHM≤ 2.0 keV FWHM≥ 77:12.0 (maximum)3.0 (typical)95 mm (nominal)GEM80P4-95 (PopTop) / GEM80-95 (Streamline)
GEM9090 % (minimum)≤ 1.20 keV FWHM≤ 2.0 keV FWHM≥ 80:12.0 (maximum)3.0 (typical)95 mm (nominal)GEM90P4-95 (PopTop) / GEM90-95 (Streamline)
GEM100100 % (minimum)≤ 1.20 keV FWHM≤ 2.1 keV FWHM≥ 83:12.0 (maximum)3.0 (typical)95 mm (nominal)GEM100P4-95 (PopTop) / GEM100-95 (Streamline)
GEM110110 % (minimum)≤ 1.30 keV FWHM≤ 2.1 keV FWHM≥ 85:12.0 (maximum)3.0 (typical)95 mm (nominal)GEM110P4-95 (PopTop) / GEM110-95 (Streamline)
GEM120120 % (minimum)≤ 1.30 keV FWHM≤ 2.1 keV FWHM≥ 86:12.0 (maximum)3.1 (typical)108 mm (nominal)GEM120P4-108 (PopTop) / GEM120-108 (Streamline)
GEM130130 % (minimum)≤ 1.30 keV FWHM≤ 2.1 keV FWHM≥ 90:12.0 (maximum)3.1 (typical)108 mm (nominal)GEM130P4-108 (PopTop) / GEM130-108 (Streamline)
GEM140140 % (minimum)≤ 1.30 keV FWHM≤ 2.2 keV FWHM≥ 90:12.0 (maximum)3.1 (typical)108 mm (nominal)GEM140P4-108 (PopTop) / GEM140-108 (Streamline)
GEM150150 % (minimum)≤ 1.30 keV FWHM≤ 2.3 keV FWHM≥ 90:12.0 (maximum)3.1 (typical)108 mm (nominal)GEM150P4-108 (PopTop) / GEM150-108 (Streamline)

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ORTEC (AMETEK)Product page Product page intro; Feature list; Options; Feature list / Options Checked 2026-10-16
  2. 2
    Publisher: ORTEC (AMETEK)Datasheet p. 1; p. 18, 'Ordering Information and Warranted Specifications'; p. 18, table 'Ordering Information and Warranted Specifications', notes 1–2; p. 18, bullet list above the ordering table … Checked 2026-10-16
Safeguards planar HPGe-detektor
ORTEC

Safeguards planar HPGe-detektor

Planar HPGe gamma detector

Detector type
Planar HPGe
Energy resolution at 122 keV
Warranted FWHM for SGD-16550P4: 550 eV at 1 kcps, 6 μs; 615 eV at 50 kcps, 1 μs
7 models
Safeguards koaxial HPGe-detektor
ORTEC

Safeguards koaxial HPGe-detektor

HPGe gamma detector

Detector type
HPGe, coaxial or semi-planar by SGD-GEM model
Energy resolution at 122 keV
FWHM maximum limits 0.61–0.80 keV at 1 kcps and optimal shaping; model-dependent
4 models
GLP plana HPGe-detektorer
ORTEC

GLP plana HPGe-detektorer

HPGe gamma detector

Energy range
Approximately 5–300 keV
Active area
Nominally 80–1,000 mm², model-dependent
Energy resolution at 122 keV
FWHM maximum limits 0.51–0.63 keV for the tabulated models
8 models
GWL brunnsdetektorer
ORTEC

GWL brunnsdetektorer

HPGe gamma detector

Detector type
P-type reverse coaxial HPGe well detector
Energy range
Approximately 10 keV–10 MeV
28 models
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.