Relative efficiency of HPGe detectors – what does 40% mean?

HPGe detectors are often specified as 20%, 40% or 100%. The figure is a comparison number defined in IEEE Std 325 – not an absolute efficiency. Here is how to read relative efficiency, resolution and peak-to-Compton ratio in a data sheet.

Step by step

How relative efficiency is measured

The measurement behind the percentage in an HPGe data sheet. Scroll to follow it.

  1. 01

    The set-up

    Relative efficiency is measured in a fixed geometry: a Co-60 point source placed 25 cm from the detector.

  2. 02

    The full-energy peak

    What counts is the full-energy peak at 1,332.5 keV – the photons that leave all their energy in the detector. The share of the source’s photons at that energy that end up in the peak is the detector’s full-energy peak efficiency.

  3. 03

    The comparison

    In the same geometry a 3″ × 3″ (76 × 76 mm) NaI(Tl) crystal has the value 1.2 × 10⁻³. Relative efficiency is the HPGe detector’s full-energy peak efficiency divided by that value – the definition in IEEE Std 325.

  4. 04

    What 40% means

    A 40% detector therefore has about 0.40 × 1.2 × 10⁻³ ≈ 4.8 × 10⁻⁴ in exactly that geometry. The figure can exceed 100% – ORTEC’s GEM series covers 10–150%. It compares detectors; the efficiency for your sample is calibrated for the actual geometry.

The definition

Relative efficiency compares the detector’s full-energy peak efficiency at 1,332.5 keV (Co-60) with that of a 3″ × 3″ (76 × 76 mm) NaI(Tl) crystal. Both refer to a point source 25 cm from the detector, and the value used for the NaI crystal is 1.2 × 10⁻³. A 40% detector therefore has a full-energy peak efficiency of about 0.40 × 1.2 × 10⁻³ ≈ 4.8 × 10⁻⁴ in exactly that geometry.

The figure compares detectors with each other – it is not the efficiency for your sample, and it can exceed 100%: ORTEC’s GEM series covers 10–150%. For quantitative measurements the efficiency is calibrated for the actual geometry, such as a Marinelli beaker or a filter, using calibration sources or computational models.

Resolution and peak-to-Compton ratio

For HPGe, resolution is stated as FWHM at 122 keV (Co-57) and 1,332 keV (Co-60). The peak-to-Compton ratio is the height of the 1,332.5 keV peak divided by the average level of the Compton continuum between 1,040 and 1,096 keV. A higher value makes it easier to see weak peaks that sit on the Compton background from stronger, higher-energy lines.

Example: two 40% detectors

ORTEC’s GEM40 (p-type) and GMX40 (n-type) have the same relative efficiency – at least 40% – but are built for different tasks:

  • GEM40: ≤ 0.87 keV at 122 keV, ≤ 1.8 keV at 1.33 MeV and peak-to-Compton ≥ 64:1. The GEM series is used from about 40 keV upwards.
  • GMX40: ≤ 1.00 keV at 122 keV, ≤ 2.0 keV at 1.33 MeV and peak-to-Compton ≥ 59:1. The thin, ion-implanted front contact (0.3 µm) lets the GMX series measure from about 5 keV.

If low-energy lines are to be measured – for example Am-241 at 59.5 keV or Pb-210 at 46.5 keV – GMX is the natural choice. For measurements above about 40 keV, the GEM40 has tighter warranted resolution figures at the same efficiency.

More efficiency or lower background?

A detector with higher relative efficiency reaches the same detection limit (MDA) in a shorter counting time, but it costs more. Background also affects the detection limit, so shielding and low-background materials can matter as much as crystal size. Tell us about the samples, nuclides and detection limit requirements and we will help you compare the options – contact us.

References

  1. IEEE Std 325-1996: IEEE Standard Test Procedures for Germanium Gamma-Ray Detectors
  2. ORTEC: GEM P-type Coaxial HPGe Radiation Detectors (product page)
  3. ORTEC: GEM Series Coaxial HPGe Detector – Product Configuration Guide, p. 18
  4. ORTEC: Gamma-X (GMX) N-type HPGe Coaxial Radiation Detectors (product page)
  5. ORTEC: GMX Series Coaxial HPGe Detector – Product Configuration Guide, p. 14

Unfamiliar term? Look it up in the glossary.

Product areas

Relevant products

GEM HPGe-detektorer
ORTEC

GEM HPGe-detektorer

HPGe detectors for gamma spectroscopy

Energy range
Approximately 40 keV and above
Relative efficiency
10–150% across the model series; higher on request
Energy resolution at 122 keV
FWHM maximum limits 0.80–1.30 keV for GEM10–GEM150; model-dependent
17 models
GAMMA-X (GMX) HPGe-detektorer
ORTEC

GAMMA-X (GMX) HPGe-detektorer

HPGe gamma detector

Energy range
Approximately 5 keV–10 MeV
Relative efficiency
10–100%; higher on request
Energy resolution at 122 keV
FWHM maximum limits 0.85–1.40 keV for GMX10–GMX100; model-dependent; optimal shaping with DSPEC-50A