Guide

Detector materials for gamma spectrometry: NaI, LaBr₃, CZT and HPGe

Energy resolution determines how close two gamma lines can be and still be told apart. This guide compares the most common detector materials using the manufacturers’ own resolution figures – and explains what the difference means when you choose an instrument.

Interactive comparison

Same source – four detector materials

1,085.8 + 1,112.1 keV
Resolution (manufacturer’s figure)≤ 7.2% FWHM at 662 keVPeak width at 662 keV in the chart: ≈ 47.7 keV
SourceORTEC RADEAGLET-R (2SG) →The value and its document are shown on the product page.

Rugged, cost-effective scintillator – common in handheld identifiers and fixed monitors.

Illustration: simulated Eu-152 spectrum (the strongest gamma lines, with energies and emission probabilities from LNHB/CEA, Table de Radionucléides). Peak widths are derived from the manufacturer’s resolution figure – scaled with √E from 662 keV for scintillators and CZT, fitted to the values at 122 keV and 1.33 MeV for HPGe. Efficiency and Compton background are simplified and identical for all materials; heights are normalised and shown on a square-root scale.

What energy resolution means

A gamma detector does not record a gamma line as a sharp line but as a bell-shaped peak. The width of the peak is given as FWHM – full width at half maximum – in keV or as a percentage of the energy. For scintillators and CZT the value is usually stated at 662 keV (Cs-137), for HPGe detectors at 122 keV (Co-57) and 1,332 keV (Co-60).

The narrower the peak, the easier it is to separate neighbouring lines and to see weak peaks on top of the background. In the chart above, the same Eu-152 source has been calculated for four detector materials. Look at the lines at 1,085.8 and 1,112.1 keV: with NaI(Tl) and LaBr₃(Ce) they merge, while CZT and HPGe show two clearly separated peaks.

The materials in brief

NaI(Tl) – sodium iodide

A scintillator that can be produced in large volumes at reasonable cost and works at room temperature. The crystal is hygroscopic and therefore always hermetically encapsulated. Its resolution is the lowest of the four materials – ORTEC states ≤ 7.2% FWHM at 662 keV for RADEAGLET-R with a NaI(Tl) crystal – but a large crystal gives high sensitivity. NaI(Tl) is common in handheld identifiers and fixed monitoring systems.

LaBr₃(Ce) – lanthanum bromide

A scintillator with clearly better resolution than NaI(Tl) at room temperature: ORTEC states ≤ 3.2% FWHM at 662 keV for RADEAGLET-R with a LaBr₃(Ce) crystal. The material contains naturally radioactive lanthanum-138, which adds a small intrinsic background to the spectrum that can be noticeable at very low activities.

CZT – cadmium zinc telluride

A semiconductor that, unlike germanium, works at room temperature. The radiation is converted directly into an electrical signal, without going through light, which gives high resolution in a compact instrument. H3D states < 1.1% FWHM at 662 keV for the H100, H400 and H420 in the H Series and < 0.8% for the plus versions. The detector volume varies between models – from 6 cm³ in the H100 to more than 77 cm³ in the H1600 – and the H Series also shows where the radiation comes from.

HPGe – high-purity germanium

Gives by far the best resolution and is the standard for laboratory measurements and demanding nuclide analysis. ORTEC warrants, for example, ≤ 0.87 keV at 122 keV and ≤ 1.8 keV at 1.33 MeV for the GEM40. The germanium crystal must be cooled to cryogenic temperature during operation, with liquid nitrogen or an electro-mechanical cooler, which makes the systems larger and more expensive.

Resolution is not everything

The right material depends on the task. A larger crystal captures more photons and shortens measuring times, so a large NaI(Tl) detector can be a better choice than a small detector with higher resolution when a source has to be found quickly. When several nuclides must be identified in a complex spectrum, or the activity of a sample determined accurately, HPGe is normally the first choice. LaBr₃(Ce) and CZT are often used in handheld identifiers, where both resolution and weight matter.

How to read resolution in a data sheet

  • Energy: check at which energy the value applies – 662 keV is most common for scintillators and CZT, 122 keV and 1,332 keV for HPGe.
  • Typical or warranted: a maximum value (≤) is a guarantee, a typical value is the normal outcome.
  • Percent and keV: the values can be converted. 7% at 662 keV corresponds to about 46 keV FWHM.
  • Conditions: resolution applies under stated conditions. The RADEAGLET-R values apply at room temperature, for example, and ORTEC’s HPGe values are measured according to ANSI/IEEE Std 325-1996 with a specified signal processor.

Want to see which instruments use a particular detector type? Use the Detector type filter in the product catalogue or start from the task in the instrument finder.

References

  1. ORTEC: RADEAGLET-R (Ruggedized) Radioisotope Identification Device, brochure 25-0312, p. 10
  2. H3D: H Series – Industrial Imaging Spectrometer, model table
  3. ORTEC: GEM Series Coaxial HPGe Detector – Product Configuration Guide, p. 18
  4. LNHB/CEA: Table de Radionucléides – Eu-152, section 5.2 Gamma Emissions

Unfamiliar term? Look it up in the glossary.

Product areas

Relevant products

RADEAGLET-R
ORTEC

RADEAGLET-R

Handheld isotope identification

Detector type
NaI(Tl) or LaBr₃(Ce), model dependent
Energy range, NaI
NaI: 15 keV–3 MeV
Spectrum channels
2,048
12 models
H Series
H3D

H Series

Gamma imaging

Detector type
CZT (CdZnTe)
Spectroscopy energy range
50–3000 keV (H100/H400/H420/H1600)
IP rating
H100/H400/H420/H1600: IP67 (excluding external media)
9 models
GEM HPGe-detektorer
ORTEC

GEM HPGe-detektorer

HPGe detectors for gamma spectroscopy

Energy range
Approximately 40 keV and above
Relative efficiency
10–150% across the model series; higher on request
Energy resolution at 122 keV
FWHM maximum limits 0.80–1.30 keV for GEM10–GEM150; model-dependent
17 models

More guides

Knowledge base
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.