A Series (A400) RIID
Handheld RIID
- Detector type
- A400: CZT (CdZnTe)
- Spectroscopy energy range
- A400: 50–3000 keV
- Weight
- A400: 1.2 kg

Bild: H3D
Gamma imaging
See where gamma radiation originates. The H Series combines spectroscopy, isotope identification and imaging in a portable system.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Detector type | CZT (CdZnTe)1 |
|---|---|
| Detector volume | H100: 6 cm³; H400/H420: >19 cm³; H1600: >77 cm³2 |
| Radiation field of view | H100/H400/H420: 4π (360°), Compton; H420: 64° × 64°, coded aperture3 |
| Energy resolution at 662 keV | <1.1% FWHM (H100, H400, H420, H1600, H1610); <0.8% FWHM (H400+, H420+, H1600+, H1610+)2 |
|---|
| Spectroscopy energy range | 50–3000 keV (H100/H400/H420/H1600)2 |
|---|---|
| Imaging energy range | H100/H400/H1600: 250–3000 keV; H420: 250–3000 keV (Compton), 50–1500 keV (coded aperture; optional extension to 3000 keV at lower efficiency)3 |
| Software | H100/H400/H420: Visualizer for post-processing5 |
|---|---|
| Nuclide library | Selected from 3,573 ENDF isotopes and user-defined; unlimited1 |
| User interface | Tablet; H100/H400/H420: 7-inch, 1920 × 10801 |
|---|---|
| Communication | H100/H400/H420: Wi-Fi, Bluetooth or wired tablet connection; Ethernet RJ45, TCP/IP5 |
| Data storage | H100/H400/H420: removable USB storage, 64 GB included1 |
| Power supply | H100/H400/H420: 100–240 V, 47–63 Hz1 |
|---|---|
| Battery life | H100 >7 h; other models >6 h2 |
| Weight (manufacturer unit) | H100 7.3 lb; H400, H420, H400+ and H420+ 7.8 lb; H1600 and H1600+ 11.7 lb; H1610 and H1610+ ~12.7 lb2 |
|---|
| IP rating | H100/H400/H420/H1600: IP67 (excluding external media)4 |
|---|---|
| Operating temperature | −20 to +50 °C; H420: coded-aperture imaging unavailable below −10 °C3 |
| Startup time | <1.5 minutes2 |
|---|
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
9 models in the series. Filter or scroll the table to compare.
| Model | Energy resolution at 662 keV2 | Imaging range2 | CZT volume2 | Weight (manufacturer unit) | Battery life | Temperature rating (manufacturer unit) | Dimensions |
|---|---|---|---|---|---|---|---|
| H100 | <1.1 % FWHM | 250–3000 keV | 6 cm³ | 7.0 lb (3.2 kg)1 | >7 hours at 23 °C; >3 hours at −20 °C or +50 °C1 | −4–122 °F2 | 24 × 8.6 × 18 cm (nominal); with the add-on Exoskeleton 37.5 × 14 × 22 cm1 |
| H400 | <1.1 % FWHM | 250–3000 keV | >19 cm³ | 7.1 lb (3.2 kg)5 | >7 hours at 23 °C; >3 hours at −20 °C or +50 °C5 | −4–122 °F2 | 24 × 9.5 × 18 cm (nominal); with the add-on Exoskeleton 37.5 × 14 × 22 cm5 |
| H400+ | <0.8 % FWHM | 250–3000 keV | >19 cm³ | 7.1 lb (3.2 kg)5 | >7 hours at 23 °C; >3 hours at −20 °C or +50 °C5 | −4–122 °F2 | 24 × 9.5 × 18 cm (nominal); with the add-on Exoskeleton 37.5 × 14 × 22 cm5 |
| H420 | <1.1 % FWHM | 50–3000 keV | >19 cm³ | 8.0 lb (3.6 kg)3 | >6 hours at 23 °C; >2 hours at −20 °C or +50 °C3 | −4–122 °F; coded-aperture imaging not available below 14 °F3 | 24 × 9.5 × 18 cm (nominal)3 |
| H420+ | <0.8 % FWHM | 50–3000 keV | >19 cm³ | 8.0 lb (3.6 kg)3 | >6 hours at 23 °C; >2 hours at −20 °C or +50 °C3 | −4–122 °F; coded-aperture imaging not available below 14 °F3 | 24 × 9.5 × 18 cm (nominal)3 |
| H1600 | <1.1 % FWHM | 250–3000 keV | >77 cm³ | 11.7 lb4 | >6 hours at 23 °C4 | −4–122 °F2 | 12.7 × 20.3 × 17.8 cm (5.0 × 8.0 × 7.0 in)4 |
| H1600+ | <0.8 % FWHM | 250–3000 keV | >77 cm³ | 11.7 lb4 | >6 hours at 23 °C4 | −4–122 °F2 | 12.7 × 20.3 × 17.8 cm (5.0 × 8.0 × 7.0 in)4 |
| H1610 | <1.1 % FWHM | 50–3000 keV | >77 cm³ | ≈12.7 lb2 | >6 hours at 23 °C4 | −4–122 °F2 | – |
| H1610+ | <0.8 % FWHM | 50–3000 keV | >77 cm³ | ≈12.7 lb2 | >6 hours at 23 °C4 | −4–122 °F2 | – |
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Handheld RIID

3D radiation mapping

Rugged integrable gamma imager

Collimated gamma imager
We help you choose the right configuration, detector and accessories for your measurement task.