CLLBC Scintillation Detectors
Dual-mode CLLBC scintillators for gamma rays and neutrons
- Energy resolution at 662 keV
- < 3.5% FWHM
- Thermal neutron peak
- Sharp peak at 3.1–3.2 MeV
- ⁶Li enrichment
- 95%

Bild: Scionix
EJ-309 liquid scintillation detectors for fast neutrons and gamma rays
Detectors with the EJ-309 liquid scintillator, which has a flash point of 144 °C and is not listed as dangerous goods. Neutrons and gamma rays are separated by pulse shape discrimination. The cell is hermetically sealed with an expansion reservoir and keeps full optical contact in every orientation. Also available boron-loaded as EJ-309:B5.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-11-26
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Scintillator | EJ-309 liquid scintillator; optionally boron-loaded EJ-309:B5 (up to 5% natural boron)1 |
|---|---|
| Active volume | Examples: 51 x 51 mm, 76 x 76 mm and 150 x 150 x 150 mm2 |
| Light sensor | Fast photomultiplier tubes (2–3 ns rise time) with transistorised voltage dividers at negative high voltage2 |
| Pulse shape discrimination | Neutron/gamma separation by PSD, just slightly inferior to EJ-3011 |
|---|---|
| Light output relative to anthracene | 75% (EJ-309:B5: 52%)1 |
| Format | Hermetically sealed cell with expansion reservoir; 100% optical contact between liquid and window in every orientation2 |
|---|---|
| Light yield | 11,500 photons/MeV (electrons); EJ-309:B5 approx. 8,0001 |
| Emission maximum | 424 nm1 |
| Density | 0.964 g/cm³ at 15 °C (EJ-309:B5: 0.963 g/cm³)1 |
| Flash point | 144 °C; not listed as dangerous goods1 |
| Decay time | Approx. 3.5 ns (short component)1 |
| Refractive index | 1.571 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
Dual-mode CLLBC scintillators for gamma rays and neutrons
CLYC:Ce scintillators for neutrons and gamma rays

76 x 76 mm (3" x 3") NaI(Tl) scintillation detector

51 x 51 mm (2" x 2") NaI(Tl) scintillation detector
We help you choose the right configuration, detector and accessories for your measurement task.