CLLBC Scintillation Detectors

Dual-mode CLLBC scintillators for gamma rays and neutrons

CLLBC gives a gamma resolution below 3.5% FWHM at 662 keV and also detects thermal neutrons thanks to lithium enriched to 95% ⁶Li, with a sharp peak at 3.1–3.2 MeV. Neutrons and gamma rays are separated by pulse shape discrimination (PSD).

Energy resolution at 662 keV
< 3.5% FWHM1
Thermal neutron peak
Sharp peak at 3.1–3.2 MeV1
⁶Li enrichment
95%1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-11-26

Manufacturer’s product page (opens in a new tab)

Already have the instrument? Request service or support →

Overview

Key features

  • Gamma rays and thermal neutrons in one crystal
  • < 3.5% FWHM at 662 keV
  • Neutron/gamma separation by PSD

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

ScintillatorCLLBC – cerium-doped caesium lithium lanthanum bromochloride1

Performance

Energy resolution at 662 keV< 3.5% FWHM1
Typical resolution by energy30 keV (¹²⁹I) 15%; 59.5 keV (²⁴¹Am) 10%; 122 keV (⁵⁷Co) 6.4%; 662 keV (¹³⁷Cs) 3.2%; 1,332 keV (⁶⁰Co) 2.3%; 2,600 keV (²²⁸Th) 1.8%1
Thermal neutron peakSharp peak at 3.1–3.2 MeV1

Electronics & signal

Pulse shape discriminationExcellent neutron/gamma discrimination using PSD1

General

⁶Li enrichment95%1
Density4.08 g/cm³1
Emission maximum420 nm1
Decay time120 ns and 500 ns (average approx. 150 ns)1
Refractive index1.901
Photoelectron yield (NaI(Tl) = 100)701
Light yieldApprox. 30,000 photons/MeV1
HygroscopicYes1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ScionixDatasheet p. 1; p. 2, table Checked 2026-11-26
ILLUSTRATION
Scionix

CLYC Scintillation Detectors

CLYC:Ce scintillators for neutrons and gamma rays

⁶Li enrichment
96%
Energy resolution at 662 keV
4.5–5.5% FWHM depending on geometry
Thermal neutron peak
3.1–3.3 MeV
13 referenced values
EJ-309 Liquid Scintillation Detectors
Scionix

EJ-309 Liquid Scintillation Detectors

EJ-309 liquid scintillation detectors for fast neutrons and gamma rays

Pulse shape discrimination
Neutron/gamma separation by PSD, just slightly inferior to EJ-301
Active volume
Examples: 51 x 51 mm, 76 x 76 mm and 150 x 150 x 150 mm
Flash point
144 °C; not listed as dangerous goods
12 referenced values
ILLUSTRATION
Scionix

LBC Scintillation Detectors

High-resolution LBC (lanthanum bromochloride) scintillation detectors

Energy resolution at 662 keV
3% FWHM (typical)
Photoelectron yield (NaI(Tl) = 100)
140
9 referenced values
76B76/3M
Scionix

76B76/3M

76 x 76 mm (3" x 3") NaI(Tl) scintillation detector

Scintillator
NaI(Tl)
Crystal dimensions
76 x 76 mm (3" x 3")
Light sensor
Photomultiplier tube, 3 in. (76 mm), bialkali photocathode
11 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.