CLLBC Scintillation Detectors
Dual-mode CLLBC scintillators for gamma rays and neutrons
- Energy resolution at 662 keV
- < 3.5% FWHM
- Thermal neutron peak
- Sharp peak at 3.1–3.2 MeV
- ⁶Li enrichment
- 95%
Illustration – product photo not available
CLYC:Ce scintillators for neutrons and gamma rays
CLYC detects both gamma rays and neutrons: thermal neutrons through lithium enriched to 96% ⁶Li and fast neutrons through the interaction with Cl-35. Neutrons and gamma rays are separated by pulse shape discrimination, and the gamma resolution is 4.5–5.5% FWHM at 662 keV depending on geometry.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-11-26
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Scintillator | CLYC – cerium-doped caesium lithium yttrium chloride (Cs₂LiYCl₆:Ce)1 |
|---|---|
| Neutron detection | Thermal neutrons through ⁶Li and fast neutrons through Cl-351 |
| Energy resolution at 662 keV | 4.5–5.5% FWHM depending on geometry1 |
|---|---|
| Thermal neutron peak | 3.1–3.3 MeV1 |
| Pulse shape discrimination | Neutrons can be discriminated from gamma rays1 |
|---|
| Emission wavelength | 275–450 nm, maximum at 370 nm1 |
|---|
| ⁶Li enrichment | 96%1 |
|---|---|
| Density | 3.31 g/cm³1 |
| Refractive index | 1.81 (at 400 nm)1 |
| Decay time | 1 ns, 50 ns and 1,000 ns1 |
| Light yield | Approx. 20,000 photons/MeV1 |
| Hygroscopic | Yes1 |
| Melting point | 640 °C1 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
Dual-mode CLLBC scintillators for gamma rays and neutrons

EJ-309 liquid scintillation detectors for fast neutrons and gamma rays

76 x 76 mm (3" x 3") NaI(Tl) scintillation detector

51 x 51 mm (2" x 2") NaI(Tl) scintillation detector
We help you choose the right configuration, detector and accessories for your measurement task.