GAMMA-X (GMX) HPGe-detektorer

HPGe gamma detector

N-type coaxial HPGe detectors for gamma spectroscopy when low-energy response and neutron resistance are important.

Energy range
Approximately 5 keV–10 MeV1
Relative efficiency
10–100%; higher on request1
Energy resolution at 122 keV
FWHM maximum limits 0.85–1.40 keV for GMX10–GMX100; model-dependent; optimal shaping with DSPEC-50A2
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Thin ion-implanted entrance contact for low-energy measurements.
  • Useful energy range from approximately 5 keV to 10 MeV.
  • N-type crystal with resistance to neutron-induced damage.

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Detector typeN-type coaxial HPGe1

Performance

Relative efficiency10–100%; higher on request1
Energy resolution at 122 keVFWHM maximum limits 0.85–1.40 keV for GMX10–GMX100; model-dependent; optimal shaping with DSPEC-50A2
Energy resolution at 1.33 MeVFWHM maximum limits 1.8–2.5 keV for GMX10–GMX100; model-dependent; optimal shaping with DSPEC-50A2
Resolution measurement conditionsTotal system resolution measured at the factory at optimal shaping time with an ORTEC DSPEC-50A (per ANSI/IEEE Std. 325-1996)2

Measuring range

Energy rangeApproximately 5 keV–10 MeV1

Electronics & signal

PreamplifierResistive feedback as standard; optional PLUS transistor-reset preamplifier (-PL) for ultra-high count rates; optional remote preamplifier (-HJ, Streamline only, not compatible with -PL or -SMN)2

Mechanical

EndcapAluminium or carbon fibre options1

Ordering & accessories

Cryostat configurationsPopTop or Streamline2
Bias controlSMART-1 (-SMN), optional1
Cooling and accessoriesCryostat and dewar or other cooling device required for operation; not included with the detector. A cryostat must be ordered with a Streamline detector.2

General

Entrance contact0.3 μm, ion implanted1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Models

12 models in the series. Filter or scroll the table to compare.

12 / 12
ModelRelative efficiency2Typical energy resolution at 5.9 keV2Energy resolution at 122 keV2Energy resolution at 1.33 MeV2Peak-to-Compton ratio2Peak shape FW.1M/FWHM2Peak shape FW.02M/FWHM2Endcap diameter2Ordering configurations2
GMX1010 % (minimum)0.60 keV FWHM; not warranted≤ 0.85 keV FWHM≤ 1.8 keV FWHM≥ 40:11.9 (maximum)2.6 (typical)70 mm (nominal)GMX10P4-70 (PopTop) / GMX10-70 (Streamline)
GMX1515 % (minimum)0.64 keV FWHM; not warranted≤ 0.88 keV FWHM≤ 1.9 keV FWHM≥ 44:11.9 (maximum)2.6 (typical)70 mm (nominal)GMX15P4-70 (PopTop) / GMX15-70 (Streamline)
GMX2020 % (minimum)0.65 keV FWHM; not warranted≤ 0.88 keV FWHM≤ 1.9 keV FWHM≥ 48:11.9 (maximum)2.8 (typical)70 mm (nominal)GMX20P4-70 (PopTop) / GMX20-70 (Streamline)
GMX2525 % (minimum)0.69 keV FWHM; not warranted≤ 0.90 keV FWHM≤ 1.9 keV FWHM≥ 48:11.9 (maximum)2.8 (typical)76 mm (nominal)GMX25P4-76 (PopTop) / GMX25-76 (Streamline)
GMX3030 % (minimum)0.72 keV FWHM; not warranted≤ 0.90 keV FWHM≤ 1.9 keV FWHM≥ 52:11.9 (maximum)2.8 (typical)76 mm (nominal)GMX30P4-76 (PopTop) / GMX30-76 (Streamline)
GMX4040 % (minimum)0.76 keV FWHM; not warranted≤ 1.00 keV FWHM≤ 2.0 keV FWHM≥ 59:12.0 (maximum)3.0 (typical)76 mm (nominal)GMX40P4-76 (PopTop) / GMX40-76 (Streamline)
GMX5050 % (minimum)0.80 keV FWHM; not warranted≤ 1.10 keV FWHM≤ 2.2 keV FWHM≥ 58:12.0 (maximum)3.0 (typical)83 mm (nominal)GMX50P4-83 (PopTop) / GMX50-83 (Streamline)
GMX6060 % (minimum)1.10 keV FWHM; not warranted≤ 1.30 keV FWHM≤ 2.3 keV FWHM≥ 56:12.0 (maximum)3.0 (typical)95 mm (nominal)GMX60P4-95 (PopTop) / GMX60-95 (Streamline)
GMX7070 % (minimum)1.10 keV FWHM; not warranted≤ 1.30 keV FWHM≤ 2.3 keV FWHM≥ 60:12.0 (maximum)3.0 (typical)95 mm (nominal)GMX70P4-95 (PopTop) / GMX70-95 (Streamline)
GMX8080 % (minimum)1.10 keV FWHM; not warranted≤ 1.30 keV FWHM≤ 2.3 keV FWHM≥ 63:12.0 (maximum)3.0 (typical)95 mm (nominal)GMX80P4-95 (PopTop) / GMX80-95 (Streamline)
GMX9090 % (minimum)1.20 keV FWHM; not warranted≤ 1.40 keV FWHM≤ 2.4 keV FWHM≥ 64:12.1 (maximum)3.1 (typical)95 mm (nominal)GMX90P4-95 (PopTop) / GMX90-95 (Streamline)
GMX100100 % (minimum)1.20 keV FWHM; not warranted≤ 1.40 keV FWHM≤ 2.5 keV FWHM≥ 64:12.2 (maximum)3.2 (typical)95 mm (nominal)GMX100P4-95 (PopTop) / GMX100-95 (Streamline)

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ORTEC (AMETEK)Product page Product page intro; Feature list; More Information; Options … Checked 2026-10-16
  2. 2
    Publisher: ORTEC (AMETEK)Datasheet p. 14, 'Ordering Information and Warranted Specifications'; p. 2, Configuration Guidelines; p. 4 capsule diagram; p. 14 'GAMMA-X Detector Options'; p. 14, 'Ordering Information and Warranted Specifications', Notes 1–2 … Checked 2026-10-16
GEM HPGe-detektorer
ORTEC

GEM HPGe-detektorer

HPGe detectors for gamma spectroscopy

Energy range
Approximately 40 keV and above
Relative efficiency
10–150% across the model series; higher on request
Energy resolution at 122 keV
FWHM maximum limits 0.80–1.30 keV for GEM10–GEM150; model-dependent
17 models
Safeguards planar HPGe-detektor
ORTEC

Safeguards planar HPGe-detektor

Planar HPGe gamma detector

Detector type
Planar HPGe
Energy resolution at 122 keV
Warranted FWHM for SGD-16550P4: 550 eV at 1 kcps, 6 μs; 615 eV at 50 kcps, 1 μs
7 models
Safeguards koaxial HPGe-detektor
ORTEC

Safeguards koaxial HPGe-detektor

HPGe gamma detector

Detector type
HPGe, coaxial or semi-planar by SGD-GEM model
Energy resolution at 122 keV
FWHM maximum limits 0.61–0.80 keV at 1 kcps and optimal shaping; model-dependent
4 models
GLP plana HPGe-detektorer
ORTEC

GLP plana HPGe-detektorer

HPGe gamma detector

Energy range
Approximately 5–300 keV
Active area
Nominally 80–1,000 mm², model-dependent
Energy resolution at 122 keV
FWHM maximum limits 0.51–0.63 keV for the tabulated models
8 models