GLP plana HPGe-detektorer

HPGe gamma detector

Small planar HPGe detectors for low-energy photons, X-rays and applications where timing performance matters.

Energy range
Approximately 5–300 keV1
Active area
Nominally 80–1,000 mm², model-dependent1
Energy resolution at 122 keV
FWHM maximum limits 0.51–0.63 keV for the tabulated models2
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Designed for low-energy measurements.
  • Suitable for measurements from approximately 5 to 300 keV.
  • Detector diameters from 10 to 36 mm, depending on model.

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Detector typePlanar HPGe (LEPS)1
Active areaNominally 80–1,000 mm², model-dependent1
Crystal thickness7–13 mm for the models in the guide2

Performance

Energy resolution at 122 keVFWHM maximum limits 0.51–0.63 keV for the tabulated models2
Energy resolution at 5.9 keVFWHM maximum limits 0.19–0.41 keV, model-dependent2

Measuring range

Energy rangeApproximately 5–300 keV1

Electronics & signal

PreamplifierOptional POF or special feedback resistor for high rates1

Mechanical

Crystal diameter10–36 mm, model-dependent1
Endcap diameter70 mm (nominal)2

Ordering & accessories

Cryostat configurationsPopTop or Streamline2
Bias controlSMART-1 (-SMN), optional1
Cooling and accessoriesCryostat and dewar or other cooling device required for operation; not included with the detector. A cryostat must be ordered with a Streamline detector.2

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Models

8 models in the series. Filter or scroll the table to compare.

8 / 8
ModelActive diameter2Active area2Crystal depth2Energy resolution at 5.9 keV2Energy resolution at 122 keV2Beryllium window thickness2Ordering configurations2
GLP-100710 mm (nominal)80 mm² (nominal)7 mm (nominal)≤ 0.19 keV FWHM≤ 0.51 keV FWHM0.1 mm (nominal); while manufacturer supplies lastGLP-1007P4 (PopTop) / GLP-1007 (Streamline)
GLP-161016 mm (nominal)200 mm² (nominal)10 mm (nominal)≤ 0.22 keV FWHM≤ 0.56 keV FWHM0.1 mm (nominal); while manufacturer supplies lastGLP-1610P4 (PopTop) / GLP-1610 (Streamline)
GLP-251025 mm (nominal)500 mm² (nominal)10 mm (nominal)≤ 0.35 keV FWHM≤ 0.58 keV FWHM0.25 mm (nominal)GLP-2510P4 (PopTop) / GLP-2510 (Streamline)
GLP-251325 mm (nominal)500 mm² (nominal)13 mm (nominal)≤ 0.32 keV FWHM≤ 0.58 keV FWHM0.25 mm (nominal)GLP-2513P4 (PopTop) / GLP-2513 (Streamline)
GLP-321032 mm (nominal)800 mm² (nominal)10 mm (nominal)≤ 0.38 keV FWHM≤ 0.61 keV FWHM0.25 mm (nominal)GLP-3210P4 (PopTop) / GLP-3210 (Streamline)
GLP-321332 mm (nominal)800 mm² (nominal)13 mm (nominal)≤ 0.36 keV FWHM≤ 0.60 keV FWHM0.25 mm (nominal)GLP-3213P4 (PopTop) / GLP-3213 (Streamline)
GLP-361036 mm (nominal)1000 mm² (nominal)10 mm (nominal)≤ 0.41 keV FWHM≤ 0.63 keV FWHM0.25 mm (nominal)GLP-3610P4 (PopTop) / GLP-3610 (Streamline)
GLP-361336 mm (nominal)1000 mm² (nominal)13 mm (nominal)≤ 0.38 keV FWHM≤ 0.62 keV FWHM0.25 mm (nominal)GLP-3613P4 (PopTop) / GLP-3613 (Streamline)

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ORTEC (AMETEK)Product page Product page intro; Feature list; Options Checked 2026-10-16
  2. 2
    Publisher: ORTEC (AMETEK)Datasheet p. 10, 'Ordering Information'; p. 1, Configuration Guidelines; p. 10, 'Ordering Information' (Endcap Diameter mm Nominal); p. 10, ordering notes … Checked 2026-10-16
GEM HPGe-detektorer
ORTEC

GEM HPGe-detektorer

HPGe detectors for gamma spectroscopy

Energy range
Approximately 40 keV and above
Relative efficiency
10–150% across the model series; higher on request
Energy resolution at 122 keV
FWHM maximum limits 0.80–1.30 keV for GEM10–GEM150; model-dependent
17 models
Safeguards planar HPGe-detektor
ORTEC

Safeguards planar HPGe-detektor

Planar HPGe gamma detector

Detector type
Planar HPGe
Energy resolution at 122 keV
Warranted FWHM for SGD-16550P4: 550 eV at 1 kcps, 6 μs; 615 eV at 50 kcps, 1 μs
7 models
Safeguards koaxial HPGe-detektor
ORTEC

Safeguards koaxial HPGe-detektor

HPGe gamma detector

Detector type
HPGe, coaxial or semi-planar by SGD-GEM model
Energy resolution at 122 keV
FWHM maximum limits 0.61–0.80 keV at 1 kcps and optimal shaping; model-dependent
4 models
GWL brunnsdetektorer
ORTEC

GWL brunnsdetektorer

HPGe gamma detector

Detector type
P-type reverse coaxial HPGe well detector
Energy range
Approximately 10 keV–10 MeV
28 models
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.