48B38/SIP-CS-E3-X

CsI(Tl) SiPM detector with temperature-compensated bias and preamplifier

Scintillation detector with a 48 x 38 mm CsI(Tl) crystal read out by a 12 x 12 mm SiPM. The bias generator and preamplifier are built in and correct the SiPM gain drift. The detector runs on 5.2–16 V and gives less than 8% FWHM at 662 keV.

Scintillator
CsI(Tl), 48 x 38 mm1
Power supply
5.2–16 V (25 mW)1
Energy resolution at 662 keV
< 8% FWHM1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-11-26

Manufacturer’s product page (opens in a new tab)

Already have the instrument? Request service or support →

Overview

Key features

  • 48 x 38 mm CsI(Tl) with SiPM
  • Built-in gain drift correction
  • 5.2–16 V, 25 mW
  • < 8% FWHM at 662 keV

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

ScintillatorCsI(Tl), 48 x 38 mm1
Light sensorSiPM, 12 x 12 mm1

Performance

Energy resolution at 662 keV< 8% FWHM1
Noise level< 10 keV1

Electronics & signal

Test bias voltage26.9 V1
Output impedance50 Ω1
Maximum output voltage−1 V1
Gain−300 mV/MeV (can be adjusted on request)1

Connectivity & interface

Electrical connectionLEMO ERA.0S.302.CLL: pin 1 power supply (5.2–16 V), pin 2 signal1

Power & battery

Power supply5.2–16 V (25 mW)1

Environment & ratings

Temperature compensationBuilt-in SiPM gain drift correction1

General

Rise time700 ns (0–100%)1
Fall time1.4 µs1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ScionixDatasheet p. 1; p. 2 Checked 2026-11-26
SiPM Scintillation Detectors
Scionix

SiPM Scintillation Detectors

Scintillation detectors with silicon photomultipliers (SiPM)

Power supply
5 V, less than 5 mA
Temperature compensation
Built-in temperature-compensated bias generator and preamplifier
Energy resolution
Comparable to photomultiplier tube readout
10 referenced values
51BM0.25/SIP-EJ444-E3-X2
Scionix

51BM0.25/SIP-EJ444-E3-X2

Thin alpha/beta SiPM detector (ZnS(Ag) + plastic scintillator)

Scintillator
ZnS(Ag) + plastic scintillator, Ø 51 x 0.25 mm (EJ-444)
Power supply
+5–16 V (3 mA)
Compatible electronics
BrightSpec Topaz-SiPM (compatible connector)
13 referenced values
76B76/3M
Scionix

76B76/3M

76 x 76 mm (3" x 3") NaI(Tl) scintillation detector

Scintillator
NaI(Tl)
Crystal dimensions
76 x 76 mm (3" x 3")
Light sensor
Photomultiplier tube, 3 in. (76 mm), bialkali photocathode
11 referenced values
51B51/2M
Scionix

51B51/2M

51 x 51 mm (2" x 2") NaI(Tl) scintillation detector

Scintillator
NaI(Tl)
Crystal dimensions
Ø 51 x 51 mm (diameter x height)
Light sensor
Photomultiplier tube, 2 in. (51 mm), bialkali photocathode
10 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.