RAM DA3-2000 Meter

Smart meter for a range of interchangeable external detectors, with internal GM detector

“Smart” meter that recognises a wide range of external detectors and selects the right readout units automatically, for area survey, contamination monitoring, preset scaling and high-dose freeze reading. RAM DA3-2000 has an internal detector with two energy-compensated GM tubes for wide-range gamma; RAM DA 2000 has no internal detector. Data are stored with date and time and downloaded to a PC.

Detector
Internal detector (DA3-2000): two energy-compensated GM tubes, ZP1201 and ZP1313 (air kerma) or ZP1202 and ZP1314 (H*(10)); DA 2000 without internal detector1
Measurement range
0.5 µSv/h – 1 Sv/h (50 µR/h – 100 R/h), internal detector1
Weight
1.5 kg including batteries and internal detectors1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-17

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Automatic detector identification and self-test
  • No recalibration when the detector is changed
  • Manually set threshold alarm; low battery, overflow and failure alarms

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

DetectorInternal detector (DA3-2000): two energy-compensated GM tubes, ZP1201 and ZP1313 (air kerma) or ZP1202 and ZP1314 (H*(10)); DA 2000 without internal detector1
Compatible external detectorsGM-40, GM-41 and GM-42 (gamma), GM-10 (β, γ, X-ray), PA-100 (alpha), IC-10 (β, γ), PM-10 (low-energy γ), PM-11 (γ); identified automatically2

Performance

Accuracy±10% of reading2

Measuring range

Measurement range0.5 µSv/h – 1 Sv/h (50 µR/h – 100 R/h), internal detector1
Energy response±30% over 50 keV–1.3 MeV; ±20% over 60 keV–1.3 MeV2
Angular dependenceLess than ±20% within ±45° of the preferred direction2

Electronics & signal

Relay outputsTwo alarm relay contacts for an external interface1

Connectivity & interface

Display range0.01 µSv/h – 1 Sv/h (1 µR/h – 100 R/h)2
Data storageData with date and time in built-in memory; download with RMV (Rotem Meter View) software1

Power & battery

Power supplyThree 1.5 V C cells: at least 150 h continuous operation with alkaline batteries and the internal detector (display lighting off), 100 h with external detectors; option: rechargeable C cells with charger or external 4.5 V DC adapter1

Mechanical

Dimensions (L × W × H)244 × 142 × 115 mm1
Weight1.5 kg including batteries and internal detectors1

Environment & ratings

Operating temperature−10 to +50 °C1

Ordering & accessories

Ordering codeBAK-1640 (RAM DA 2000), BAK-1670 (RAM DA 3-2000, µSv/h), BAK-1660 (RAM DA 3-2000, mR/h)1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: Rotem IndustriesDatasheet TDS p. 1; TDS; TDS, Ordering Info; p. 3, Technical Data – Power Source … Checked 2026-10-16
  2. 2
    Publisher: Rotem IndustriesProduct page Product page – Technical Information; Product page – Product Specifications Checked 2026-10-16
GM-40 Detector
Rotem

GM-40 Detector

GM gamma detector for very high fields, 250 µSv/h – 10 Sv/h

Detector
GM tube ZP1301
Measurement range
250 µSv/h – 10 Sv/h (25 mR/h – 1,000 R/h)
Energy range
50 keV – 1.3 MeV
8 referenced values
GM-41 Detector
Rotem

GM-41 Detector

GM gamma detector for high fields, e.g. in cyclotron bunkers and hot cells

Detector
GM tube ZP-1313 or equivalent
Measurement range
50 µSv/h – 1 Sv/h (5 mR/h – 100 R/h)
Energy range
50 keV – 1.3 MeV
10 referenced values
GM-42 Detector
Rotem

GM-42 Detector

Sensitive low-range GM gamma detector for clean rooms and radiopharmacies

Detector
GM tube ZP-1201 or equivalent
Measurement range
0.5 µSv/h – 10 mSv/h (50 µR/h – 1 R/h)
Energy range
50 keV – 1.3 MeV
9 referenced values
GM-10 Detector
Rotem

GM-10 Detector

Pancake GM probe for alpha, beta and gamma surface contamination

Detector
Pancake GM tube LND 73118 or equivalent; mica window 1.5–2.0 mg/cm² behind a fine stainless-steel wire mesh
Count-rate range
0–42,000 cps
Weight
370 g (0.8 lb)
8 referenced values
PA-100 Detector
Rotem

PA-100 Detector

Air proportional alpha detector with a 100 cm² sensitive area

Detector
Air proportional alpha detector
Weight
1 kg
9 referenced values
IC-10 Detector
Rotem

IC-10 Detector

Rugged ion chamber detector for gamma, X-ray and beta radiation

Detector
Ion chamber, 500 cc; wall and cover 300 mg/cm² (tissue equivalent); end window 7 mg/cm²
Measurement range
1 µSv/h – 250 mSv/h (0.1 mR/h – 25 R/h)
Weight
470 g
10 referenced values
PM-10 Detector
Rotem

PM-10 Detector

Thin NaI(Tl) probe for low-energy gamma and X-rays, optimised for I-125

Detector
NaI(Tl), 2 in diameter × 0.04 in; 1 mil (0.001 in) aluminium window
Energy range
Low-energy gamma, 10–80 keV
Count-rate range
0–50,000 cps
8 referenced values
PM-11 Detector
Rotem

PM-11 Detector

Sensitive NaI(Tl) gamma probe, Rotem’s primary effluent detector in WebiSmarts

Detector
NaI(Tl), 5.1 × 5.1 cm (2 × 2 in); magnetically shielded 5.1 cm photomultiplier
Energy range
Gamma above 50 keV; recommended 30 keV – 3.0 MeV
Count-rate range
0–50,000 cps
8 referenced values