IC-10 Detector

Rugged ion chamber detector for gamma, X-ray and beta radiation

Rugged ion chamber detector for accurate and highly stable field measurements of exposure rates from gamma, X-ray and beta radiation over a wide range. It is used with the RAM DA-2000 or in the WebiSmarts system and is also available with a 6 MeV energy sleeve.

Detector
Ion chamber, 500 cc; wall and cover 300 mg/cm² (tissue equivalent); end window 7 mg/cm²1
Measurement range
1 µSv/h – 250 mSv/h (0.1 mR/h – 25 R/h)1
Weight
470 g1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-17

Manufacturer’s product page (opens in a new tab)

Overview

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

DetectorIon chamber, 500 cc; wall and cover 300 mg/cm² (tissue equivalent); end window 7 mg/cm²1

Performance

Accuracy±15% of reading within the measuring range1

Measuring range

Measurement range1 µSv/h – 250 mSv/h (0.1 mR/h – 25 R/h)1
Sensitivity0.04 pA per mR/h (gamma)1
Energy responseGamma: better than ±20% at 20 keV–1.2 MeV (Cs-137)1
Angular dependenceLess than ±5% within ±120° of the front direction (Cs-137)1

Connectivity & interface

Display range0.1 µSv/h – 250 mSv/h1

Mechanical

DimensionsChamber: 11.7 cm long, ø8.9 cm; handle: 14.3 cm long, ø5.2 cm1
Weight470 g1

Environment & ratings

Operating temperature−10 to +50 °C1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: Rotem IndustriesDatasheet TDS, Technical Data; Specifications – Dimensions; Specifications – Temperature Range; Specifications – Weight Checked 2026-10-16
RAM DA3-2000 Meter
Rotem

RAM DA3-2000 Meter

Smart meter for a range of interchangeable external detectors, with internal GM detector

Detector
Internal detector (DA3-2000): two energy-compensated GM tubes, ZP1201 and ZP1313 (air kerma) or ZP1202 and ZP1314 (H*(10)); DA 2000 without internal detector
Measurement range
0.5 µSv/h – 1 Sv/h (50 µR/h – 100 R/h), internal detector
Weight
1.5 kg including batteries and internal detectors
14 referenced values
WebiSmarts System
Rotem

WebiSmarts System

Web-based area and stack monitoring system with DPU-3 monitors

Primary function
Web software that receives online data from DPU-3 area monitors and shows them on colour-coded maps
Functions
One year of history per point; isotope identification of effluent; over 20 maps; real-time background subtraction; dose and dose rate at the same time
System components
Web software, DPU-3 monitors (four detectors each) and a stack monitor; maps, points and history are stored on a server and accessed with a browser
4 referenced values
PM-33 Detector
Rotem

PM-33 Detector

Highly sensitive 3 × 3 inch NaI(Tl) probe for environmental gamma monitoring

Energy range
Gamma above 50 keV
Measurement range
Stand-alone: background to 10 µSv/h; as environmental monitor with the TeleMeter: background to 15 mSv/h
Count-rate range
0–50,000 cps
9 referenced values
PM-11 Detector
Rotem

PM-11 Detector

Sensitive NaI(Tl) gamma probe, Rotem’s primary effluent detector in WebiSmarts

Detector
NaI(Tl), 5.1 × 5.1 cm (2 × 2 in); magnetically shielded 5.1 cm photomultiplier
Energy range
Gamma above 50 keV; recommended 30 keV – 3.0 MeV
Count-rate range
0–50,000 cps
8 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.