ORTEC C Series
Annular partially depleted surface-barrier detectors
- Detector type
- Annular partially depleted silicon surface barrier
Bild: ORTEC
Low-background silicon detectors for alpha spectroscopy
ORTEC ULTRA-AS comprises 8 models (low-background silicon detectors for alpha spectroscopy). The models differ in active area, depletion depth and warranted resolution, among other things – compare them in the model table.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Detector type | Ion-implanted silicon, partially depleted1 |
|---|---|
| Active depth | ≈100 µm2 |
| Resolution measurement conditions | ²⁴¹Am 5.486 MeV alphas; standard ORTEC electronics; 1 µs shaping time constants3 |
| Resolution measurement and warranty temperature | 21 ±1 °C1 |
| Bias polarity | Positive3 |
| Resolution reference geometry | Thin-window point source; distance at least 2 times the detector diameter4 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
8 models in the series. Filter or scroll the table to compare.
| Model | Minimum active area2 | Guaranteed maximum alpha resolution2 | Ordering mounts2 | Alpha Suite compatibility2 | Installation and spare packages |
|---|---|---|---|---|---|
| U-019-300-AS | ≥300 mm² | ≤19 keV FWHM | B, C, T (closed back) | ULTRA-AS in B mount is guaranteed compatible; C/T mounts require an approved special | B, C or T mount; ENS-U300: B mount, installed and system tested with an Alpha Suite spectrometer; ENS8U300: pack of eight spare detectors (not installed or system tested), only with an ALPHA-ENSEMBLE-8 system2 |
| U-020-450-AS | ≥450 mm² | ≤20 keV FWHM | B, C, T (closed back) | ULTRA-AS in B mount is guaranteed compatible; C/T mounts require an approved special | B, C or T mount; ENS-U450: B mount, installed and system tested with an Alpha Suite spectrometer; ENS8U450: pack of eight spare detectors (not installed or system tested), only with an ALPHA-ENSEMBLE-8 system2 |
| U-020-490-AS | ≥490 mm² | ≤20 keV FWHM | B, C, T (closed back) | ULTRA-AS in B mount is guaranteed compatible; C/T mounts require an approved special | B, C or T mount; ENS-U490: B mount, installed and system tested with an Alpha Suite spectrometer; ENS8U490: pack of eight spare detectors (not installed or system tested), only with an ALPHA-ENSEMBLE-8 system2 |
| U-024-600-AS | ≥600 mm² | ≤24 keV FWHM | B, C, T (closed back) | ULTRA-AS in B mount is guaranteed compatible; C/T mounts require an approved special | B, C or T mount; ENS-U600: B mount, installed and system tested with an Alpha Suite spectrometer; ENS8U600: pack of eight spare detectors (not installed or system tested), only with an ALPHA-ENSEMBLE-8 system2 |
| U-029-900-AS | ≥900 mm² | ≤29 keV FWHM | B, C | ULTRA-AS in B mount is guaranteed compatible; C mount requires an approved special | B or C mount; ENS-U900: B mount, installed and system tested with an Alpha Suite spectrometer; ENS8U900: pack of eight spare detectors (not installed or system tested), only with an ALPHA-ENSEMBLE-8 system2 |
| U-037-1200-AS | ≥1200 mm² | ≤37 keV FWHM | B, C | ULTRA-AS in B mount is guaranteed compatible; C mount requires an approved special | B or C mount; ENS-U1200: B mount, installed and system tested with an Alpha Suite spectrometer; ENS8U1200: pack of eight spare detectors (not installed or system tested), only with an ALPHA-ENSEMBLE-8 system2 |
| U-053-2000-AS | ≥2000 mm² | ≤53 keV FWHM | B, C | ULTRA-AS in B mount is guaranteed compatible; C mount requires an approved special | B or C mount; ENS-U2000: B mount, installed and system tested with an Alpha Suite spectrometer3 |
| U-064-3000-AS | ≥3000 mm² | ≤64 keV FWHM | B, C | ULTRA-AS in B mount is guaranteed compatible; C mount requires an approved special | B or C mount; ENS-U3000: B mount, installed and system tested with an Alpha Suite spectrometer3 |
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Annular partially depleted surface-barrier detectors

Totally depleted silicon surface-barrier detectors

Partially depleted silicon surface-barrier detectors

We help you choose the right configuration, detector and accessories for your measurement task.