ORTEC C Series
Annular partially depleted surface-barrier detectors
- Detector type
- Annular partially depleted silicon surface barrier

Bild: ORTEC
Partially depleted silicon surface-barrier detectors
ORTEC A Series comprises 29 models (partially depleted silicon surface-barrier detectors). The models differ in active area, depletion depth and warranted resolution, among other things – compare them in the model table.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Detector type | Partially depleted silicon surface barrier1 |
|---|---|
| Resolution measurement conditions | ²⁴¹Am, 5.486 MeV alphas; standard ORTEC electronics; 0.5 µs shaping time constants1 |
| Measurement and warranty temperature | 21 ±1 °C2 |
| Resolution reference geometry | Thin-window point source; distance at least 2 times the detector diameter2 |
| Mount options per product guide | A/B/C3 |
| Alpha Suite compatibility | Not guaranteed compatible; requires an approved special for compatibility and warranted specifications3 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
29 models in the series. Filter or scroll the table to compare.
| Model | Minimum active area3 | Guaranteed maximum alpha resolution4 | Minimum depletion depth4 | Ordering condition1 |
|---|---|---|---|---|
| A-013-025-1000 | ≥25 mm² | ≤13 keV FWHM | ≥1000 µm | – |
| A-014-025-1000 | ≥25 mm² | ≤14 keV FWHM | ≥1000 µm | – |
| A-014-025-1500 | ≥25 mm² | ≤14 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-014-050-1000 | ≥50 mm² | ≤14 keV FWHM | ≥1000 µm | – |
| A-014-100-1000 | ≥100 mm² | ≤14 keV FWHM | ≥1000 µm | – |
| A-015-050-1000 | ≥50 mm² | ≤15 keV FWHM | ≥1000 µm | – |
| A-015-100-1000 | ≥100 mm² | ≤15 keV FWHM | ≥1000 µm | – |
| A-015-150-1000 | ≥150 mm² | ≤15 keV FWHM | ≥1000 µm | – |
| A-016-025-1500 | ≥25 mm² | ≤16 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-016-025-2000 | ≥25 mm² | ≤16 keV FWHM | ≥2000 µm | This depth requires special order according to manufacturer |
| A-016-050-1500 | ≥50 mm² | ≤16 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-016-150-1000 | ≥150 mm² | ≤16 keV FWHM | ≥1000 µm | – |
| A-017-050-2000 | ≥50 mm² | ≤17 keV FWHM | ≥2000 µm | This depth requires special order according to manufacturer |
| A-017-100-1500 | ≥100 mm² | ≤17 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-017-300-1000 | ≥300 mm² | ≤17 keV FWHM | ≥1000 µm | – |
| A-018-050-1500 | ≥50 mm² | ≤18 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-018-100-2000 | ≥100 mm² | ≤18 keV FWHM | ≥2000 µm | This depth requires special order according to manufacturer |
| A-018-150-1500 | ≥150 mm² | ≤18 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-018-300-1000 | ≥300 mm² | ≤18 keV FWHM | ≥1000 µm | – |
| A-018-450-1000 | ≥450 mm² | ≤18 keV FWHM | ≥1000 µm | – |
| A-019-100-1500 | ≥100 mm² | ≤19 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-019-150-2000 | ≥150 mm² | ≤19 keV FWHM | ≥2000 µm | This depth requires special order according to manufacturer |
| A-019-450-1000 | ≥450 mm² | ≤19 keV FWHM | ≥1000 µm | – |
| A-020-150-1500 | ≥150 mm² | ≤20 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-020-300-1500 | ≥300 mm² | ≤20 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-021-450-1500 | ≥450 mm² | ≤21 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-022-300-2000 | ≥300 mm² | ≤22 keV FWHM | ≥2000 µm | This depth requires special order according to manufacturer |
| A-023-300-1500 | ≥300 mm² | ≤23 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
| A-026-450-1500 | ≥450 mm² | ≤26 keV FWHM | ≥1500 µm | This depth requires special order according to manufacturer |
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Annular partially depleted surface-barrier detectors

Totally depleted silicon surface-barrier detectors

Low-background silicon detectors for alpha spectroscopy
We help you choose the right configuration, detector and accessories for your measurement task.