PM-33 Detector

Highly sensitive 3 × 3 inch NaI(Tl) probe for environmental gamma monitoring

Gamma scintillation probe with a hermetically sealed 3 × 3 inch NaI(Tl) crystal and photomultiplier. It can be used as an external detector for the RAM DA-2000, and together with the TeleMeter it becomes an extremely sensitive portable environmental monitor measuring from background to 15 mSv/h.

Energy range
Gamma above 50 keV1
Measurement range
Stand-alone: background to 10 µSv/h; as environmental monitor with the TeleMeter: background to 15 mSv/h1
Count-rate range
0–50,000 cps1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-17

Manufacturer’s product page (opens in a new tab)

Overview

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

DetectorNaI(Tl), 3 × 3 in1

Measuring range

Energy rangeGamma above 50 keV1
Sensitivity4,500 cps per µSv/h (¹³⁷Cs)2
Measurement rangeStand-alone: background to 10 µSv/h; as environmental monitor with the TeleMeter: background to 15 mSv/h1
Count-rate range0–50,000 cps1

Mechanical

DimensionsLength 365 mm, width 105 mm1
Weight2.95 kg (6.5 lb)1

Environment & ratings

Operating temperature−10 to +50 °C1

Ordering & accessories

Ordering codeBAK-15151

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: Rotem IndustriesDatasheet TDS Checked 2026-10-16
  2. 2
    Publisher: Rotem IndustriesProduct page Product page – Technical Information Checked 2026-10-16
EnviSmarts
Rotem

EnviSmarts

Indoor and outdoor environmental monitoring with stand-alone monitors and online software

Measurement range
PM-33 monitor: background to 15 mSv/h (1.5 R/h), the detector up to 10 µSv/h and then the TeleMeter’s GM tube; RP-33 monitor: background to 1 Sv/h (100 R/h) with the RAM R-200’s GM tubes from 10 µSv/h
3 referenced values
RAM DA3-2000 Meter
Rotem

RAM DA3-2000 Meter

Smart meter for a range of interchangeable external detectors, with internal GM detector

Detector
Internal detector (DA3-2000): two energy-compensated GM tubes, ZP1201 and ZP1313 (air kerma) or ZP1202 and ZP1314 (H*(10)); DA 2000 without internal detector
Measurement range
0.5 µSv/h – 1 Sv/h (50 µR/h – 100 R/h), internal detector
Weight
1.5 kg including batteries and internal detectors
14 referenced values
PM-11 Detector
Rotem

PM-11 Detector

Sensitive NaI(Tl) gamma probe, Rotem’s primary effluent detector in WebiSmarts

Detector
NaI(Tl), 5.1 × 5.1 cm (2 × 2 in); magnetically shielded 5.1 cm photomultiplier
Energy range
Gamma above 50 keV; recommended 30 keV – 3.0 MeV
Count-rate range
0–50,000 cps
8 referenced values
PM-10 Detector
Rotem

PM-10 Detector

Thin NaI(Tl) probe for low-energy gamma and X-rays, optimised for I-125

Detector
NaI(Tl), 2 in diameter × 0.04 in; 1 mil (0.001 in) aluminium window
Energy range
Low-energy gamma, 10–80 keV
Count-rate range
0–50,000 cps
8 referenced values