WDMS NT-VK

Monitoring and disposal of radioactive wastewater

System that collects and monitors radioactive wastewater, mainly from diagnostic and therapeutic procedures with radioactive substances, and releases it only after its activity has dropped below a defined value. The water is stored in decay tanks and measured in Marinelli geometry with a NaI(Tl) detector in a lead well; a local PLC is commanded from software on a remote PC.

Detector type
NaI(Tl) 2 × 2 in with MCA2
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-17

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Redundant level sensors and pumps
  • Flooding protection with a safety well
  • Customised to the installation

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

UseCollection and monitoring of radioactive wastewater, mainly from diagnostic and therapeutic procedures with radioactive substances; release only once the activity is below a defined value1
Detector typeNaI(Tl) 2 × 2 in with MCA2
Sample geometry1-litre Marinelli beaker with connections to the system2
Lead shielding5 cm lead well2
Main componentsImhoff purification tanks, sorting pumps, decay tanks, sampling system, release pumps, and level and pump sensors with a safety flooding well1
ControlLocal PLC commanded by software on a remote PC: automatic cycles, measurement parameters, alarm and release archives and system status1
RedundancyCritical elements are redundant: a second Imhoff tank takes over when the first is clogged or serviced, and the sorting pumps work alternately2
CustomisationNumber and type of decay tanks defined by the installation2

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ELSE NUCLEARProduct page Product page Checked 2026-10-16
  2. 2
    Publisher: ELSE NUCLEARDatasheet p. 2, Technical Specifications; p. 2, Safety Devices; p. 2, System layout Checked 2026-10-16
LEM vätskeeffluentmonitor
ELSE NUCLEAR

LEM vätskeeffluentmonitor

Liquid effluent monitor

Detector type
2 × 2 inch NaI(Tl)
Energy range
30 keV–2 MeV
Maximum frame weight
500 kg
12 referenced values
FREE-AND 16
ELSE NUCLEAR

FREE-AND 16

Spectrometric station with 16 NaI(Tl) for waste characterisation

Detector type
16 NaI(Tl) 3 × 3 in crystals with PMT and MCA, eight on each side in two arrays
9 referenced values
AURAS 3000
ELSE NUCLEAR

AURAS 3000

Free-release assay of large waste containers with up to four HPGe

Energy range
HPGe: 30 keV–7 MeV
Energy resolution at 1.33 MeV
HPGe: ≤2.3 keV FWHM for ⁶⁰Co (1332.5 keV)
Relative efficiency
15–150% (nitrogen-cooled), 50–150% (electro-cooled)
13 referenced values
ARCHIMEDES III
ELSE NUCLEAR

ARCHIMEDES III

Portal monitor for excavation materials in boxes and containers

Detector type
Four plastic scintillators 100 × 50 × 5 cm
Active area
5000 cm² per detector; two pairs mounted on the sides of the portal
Energy range
35 keV–2 MeV
8 referenced values