LEM vätskeeffluentmonitor

Liquid effluent monitor

A system that samples liquid effluents and analyzes radioactivity using gamma spectrometry.

Detector type
2 × 2 inch NaI(Tl)2
Energy range
30 keV–2 MeV2
Maximum frame weight
500 kg2

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Sampling of liquid effluents.
  • Spectrometric analysis in Marinelli geometry.
  • Managed with ELSE NUCLEAR software.

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Detector type2 × 2 inch NaI(Tl)2
Sample geometry1 L plexiglass Marinelli1

Performance

Efficiency in measurement geometry, Cs-137Over 1.5%2

Measuring range

Energy range30 keV–2 MeV2
Response in measurement geometry, Cs-137Over 12 cps/kBq2

Power & battery

Power-failure actionAlarm and interlock activate2

Mechanical

Frame dimensions (W × D × H)80 × 70 × 150 cm2
Maximum frame weight500 kg2

General

Measurement functionsSpectrometric analysis of liquid effluents1
Result unitsBq/L, cps or cpm2
Lead shielding50 mm standard; 100 mm optional2
Sampling system working pressure10 bar when sampling equipment is supplied with the system2

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ELSE NUCLEARProduct page product page Checked 2026-10-16
  2. 2
    Publisher: ELSE NUCLEARDatasheet p. 2 Checked 2026-10-16
FREE-AND 16
ELSE NUCLEAR

FREE-AND 16

Spectrometric station with 16 NaI(Tl) for waste characterisation

Detector type
16 NaI(Tl) 3 × 3 in crystals with PMT and MCA, eight on each side in two arrays
9 referenced values
AURAS 3000
ELSE NUCLEAR

AURAS 3000

Free-release assay of large waste containers with up to four HPGe

Energy range
HPGe: 30 keV–7 MeV
Energy resolution at 1.33 MeV
HPGe: ≤2.3 keV FWHM for ⁶⁰Co (1332.5 keV)
Relative efficiency
15–150% (nitrogen-cooled), 50–150% (electro-cooled)
13 referenced values
ARCHIMEDES III
ELSE NUCLEAR

ARCHIMEDES III

Portal monitor for excavation materials in boxes and containers

Detector type
Four plastic scintillators 100 × 50 × 5 cm
Active area
5000 cm² per detector; two pairs mounted on the sides of the portal
Energy range
35 keV–2 MeV
8 referenced values
ADAMOS
ELSE NUCLEAR

ADAMOS

Automated segmented gamma scanning of waste drums

Energy range
HPGe: 30 keV–7 MeV
Energy resolution at 1.33 MeV
HPGe: ≤2.3 keV FWHM for ⁶⁰Co (1332.5 keV)
Relative efficiency
15–150% (nitrogen-cooled), 50–150% (electro-cooled)
13 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.