LBC Scintillation Detectors
High-resolution LBC (lanthanum bromochloride) scintillation detectors
- Energy resolution at 662 keV
- 3% FWHM (typical)
- Photoelectron yield (NaI(Tl) = 100)
- 140

Bild: Scionix
High-resolution, low-background CeBr₃ scintillation detectors
CeBr₃ is an alternative to NaI(Tl) for high-resolution gamma spectrometry: typically 4% FWHM at 662 keV, and better than NaI(Tl) above 200 keV. Unlike lanthanum halides, CeBr₃ has no La-138 background. Crystals are available up to Ø 127 mm and 152 mm high.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-11-26
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Scintillator | CeBr₃ (cerium bromide)1 |
|---|
| Energy resolution at 662 keV | 4% FWHM (typical)1 |
|---|---|
| Typical resolution by energy | 30 keV (¹²⁹I) 20%; 59.5 keV (²⁴¹Am) 13%; 122 keV (⁵⁷Co) 8%; 662 keV (¹³⁷Cs) 4%; 1,332 keV (⁶⁰Co) 3%; 2,600 keV (²²⁸Th) 2.5%1 |
| Compared with NaI(Tl) | Better resolution than NaI(Tl) above 200 keV1 |
| Maximum crystal size | Ø 127 mm, 152 mm high1 |
|---|
| Intrinsic background | No La-138 background; peaks between 1,500 and 2,200 keV from Ac-227, typically 0.002 c/s per cm³1 |
|---|
| Decay time | 18–25 ns, size dependent (typical)1 |
|---|---|
| Emission maximum | 370 nm1 |
| Refractive index | 2.09 (at 380 nm)1 |
| Light yield | Approx. 60,000 photons/MeV1 |
| Hygroscopic | Yes1 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
High-resolution LBC (lanthanum bromochloride) scintillation detectors

High-resolution SrI₂(Eu) scintillation detectors

Scintillation detectors with silicon photomultipliers (SiPM)

76 x 76 mm (3" x 3") NaI(Tl) scintillation detector
We help you choose the right configuration, detector and accessories for your measurement task.