RG-40 Detector

GM probe for high gamma fields, up to 100 Sv/h, for the RAM R-200

GM-based probe for detecting high gamma radiation. Together with the RAM R-200 it gives measurements over a very wide range, up to 100 Sv/h.

Detector
GM tube 4G60M1
Measurement range
1 mSv/h – 100 Sv/h (0.1–10,000 R/h)1
Weight
130 g (0.3 lb)1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-17

Manufacturer’s product page (opens in a new tab)

Overview

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

DetectorGM tube 4G60M1

Performance

Accuracy±10% of reading1

Measuring range

Measurement range1 mSv/h – 100 Sv/h (0.1–10,000 R/h)1
Energy response±35% from 60 keV to 2.0 MeV1
Sensitivity77 cps per R/h1

Mechanical

Dimensions145 × ø30 mm1
Weight130 g (0.3 lb)1

Environment & ratings

Operating temperature−30 to +60 °C1
IP ratingIP67 (splash-proof aluminium casing)1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: Rotem IndustriesDatasheet RAM R-200 TDS p. 3, External Detectors – RG-40 Checked 2026-10-16
RAM R-200 Meter
Rotem

RAM R-200 Meter

Rugged multipurpose survey meter for wide-range gamma and contamination

Detector
Two energy-compensated GM tubes: ZP-1202 and ZP1314 (or equivalent)
Measurement range
Internal detectors: 0.1 µSv/h – 1 Sv/h (10 µR/h – 100 R/h); with external probes gamma up to 100 Sv/h and alpha/beta 0–1 Mcpm
Weight
460 g including battery
15 referenced values
PM-33 Detector
Rotem

PM-33 Detector

Highly sensitive 3 × 3 inch NaI(Tl) probe for environmental gamma monitoring

Energy range
Gamma above 50 keV
Measurement range
Stand-alone: background to 10 µSv/h; as environmental monitor with the TeleMeter: background to 15 mSv/h
Count-rate range
0–50,000 cps
9 referenced values
PM-11 Detector
Rotem

PM-11 Detector

Sensitive NaI(Tl) gamma probe, Rotem’s primary effluent detector in WebiSmarts

Detector
NaI(Tl), 5.1 × 5.1 cm (2 × 2 in); magnetically shielded 5.1 cm photomultiplier
Energy range
Gamma above 50 keV; recommended 30 keV – 3.0 MeV
Count-rate range
0–50,000 cps
8 referenced values
PM-10 Detector
Rotem

PM-10 Detector

Thin NaI(Tl) probe for low-energy gamma and X-rays, optimised for I-125

Detector
NaI(Tl), 2 in diameter × 0.04 in; 1 mil (0.001 in) aluminium window
Energy range
Low-energy gamma, 10–80 keV
Count-rate range
0–50,000 cps
8 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.