RADEAGLET-R
Handheld isotope identification
- Detector type
- NaI(Tl) or LaBr₃(Ce), model dependent
- Energy range, NaI
- NaI: 15 keV–3 MeV
- Spectrum channels
- 2,048

Bild: ORTEC
Alpha/beta probe for the RADEAGLET-R
Handheld scintillation probe that extends the RADEAGLET-R to alpha and beta surface contamination measurements. It connects by USB, is powered by the instrument and can be attached or detached during operation; pulse shape discrimination separates alpha from beta.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-17
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Radiation type | Alpha and beta1 |
|---|---|
| Detector type | Alpha/beta-sensitive scintillator with photomultiplier (ZnS:Ag)1 |
| Window | Mylar foil with honeycomb protective grid1 |
| Alpha/beta separation | Pulse shape discrimination (PSD) in a multichannel analyser with 4096 energy channels × 4096 width channels, a non-linear parametric window and Haykin adaptive optimisation1 |
| Count-rate range | Up to 25 kcps alpha and 20 kcps beta1 |
|---|
| Communication | List mode; USB interrupt-mode transfer1 |
|---|
| Power supply | 5 V, USB-powered by the RADEAGLET-R1 |
|---|
| Detector dimensions | ø50 mm (2.0 in)1 |
|---|---|
| Weight | 300 g (0.66 lb)1 |
| Dimensions | 169 × 68 mm (6.7 × 2.7 in); handle ø50 mm (2.0 in)1 |
| Operating temperature | −30 °C to +70 °C1 |
|---|---|
| Operating time | 8 hours for the RADEAGLET-R with the probe connected1 |
| Part number | RTR-IP-2AB1 |
|---|
| Processor | High-performance, low-power FPGA subsystem coupled to a DSP1 |
|---|---|
| Compatibility | RADEAGLET-R; can be attached or detached during operation and is auto-calibrated1 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Handheld isotope identification

Air proportional alpha detector with a 100 cm² sensitive area

Pancake GM probe for alpha, beta and gamma surface contamination

Alpha and beta probes for the SaphyRAD contamination meter