BSS
Bonner sphere spectrometer for neutron spectra
- Detector type
- ³He proportional counters for thermal neutrons, each in a spherical HDPE moderator
- Energy range
- 1 meV to <100 MeV (standard set); up to 10 GeV with extension set

Bild: ELSE NUCLEAR
Single-sphere neutron spectrometer instead of a full Bonner sphere set
Active neutron spectrometer in which 32 silicon detectors in six layers inside one moderating sphere reproduce the spectrometric performance of a six-sphere Bonner set. It measures the neutron spectrum from thermal up to 10 GeV (20 MeV in the LITE version), has built-in electronics and can show the spectrum and H*(10) in real time. It tolerates a high gamma background.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-17
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Detector type | 32 active ⁶LiF-covered silicon detectors in six concentric layers inside the moderator1 |
|---|
| Energy range | 0.025 eV–10 GeV (standard); 0.025 eV–20 MeV (LITE)2 |
|---|---|
| Typical neutron sensitivity | Approx. 0.3 cps/(µSv/h)2 |
| Gamma sensitivity | Completely insensitive up to 50 mSv/h continuous gamma background2 |
| Angular dependence | Max. ±2% for a ²⁴¹Am-Be source2 |
| Maximum H*(10) per burst in pulsed fields | 1–5 µSv depending on the neutron spectrum2 |
|---|
| Software | Built-in unfolding algorithm for spectrum and H*(10) on-line; raw data can be saved for off-line analysis1 |
|---|
| Power supply | 230 V ±10%, 50 Hz2 |
|---|
| Options | LITE version; traceable calibration (1 dose-rate point, Am-Be) and warranty extension (24 months) on request2 |
|---|
| Maximum counting rate | 10⁵ cps per acquisition channel2 |
|---|
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Bonner sphere spectrometer for neutron spectra

Very high-efficiency neutron spectrometer, from thermal to 10 GeV

Neutron rem counter for pulsed fields