LUPIN BF3

Neutron rem counter for pulsed fields

Monitoring unit for neutron H*(10) with unusually good performance in pulsed radiation fields, for example around accelerators. The rem counter has a BF₃ proportional counter in a cylindrical moderator and built-in electronics that calculate the H*(10) rate every second and send it to a SATURN ratemeter. The LITE version is lighter and measures up to 20 MeV.

Detector type
BF₃ proportional counter in a cylindrical moderator, with built-in power supply, signal management and control electronics1
Energy range
0.025 eV–10 GeV (standard); 0.025 eV–20 MeV (LITE)2
Weight
18 kg; 15 kg (LITE)2
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-17

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Energy response fitting the ICRP74 conversion curve
  • Unaffected by signal pile-up
  • Optional ultra-fast alarm (50 ms) – doubles as a beam loss monitor

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Detector typeBF₃ proportional counter in a cylindrical moderator, with built-in power supply, signal management and control electronics1

Measuring range

Energy range0.025 eV–10 GeV (standard); 0.025 eV–20 MeV (LITE)2
Dose-rate rangeH*(10): 10 nSv/h–100 mSv/h2
Neutron sensitivity0.6 cps/(µSv/h)2
Gamma sensitivity<0.5 µSv/h at 50 mSv/h (662 keV)2
Angular dependence<20%2
Energy responseFits the ICRP74 conversion curve; unaffected by signal pile-up1

Electronics & signal

Maximum H*(10) per burst in pulsed fields2 µSv with underestimation ≤10%2

Connectivity & interface

CommunicationRS232 (default), RS485/422 for long distances, Ethernet up to 1 km2

Power & battery

Power supplyBuilt-in power supply with signal management and control electronics2

Mechanical

DimensionsØ250 × 425 mm2
Weight18 kg; 15 kg (LITE)2

Environment & ratings

Operating temperature0–50 °C2
IP ratingIP54 version available as an accessory2

Ordering & accessories

OptionsLITE version, radiation-sensitive electronics in a separate rack, ultra-fast response (alarm in 50 ms), battery-operated version; on request traceable calibration (Am-Be), trolley, IP54 configuration and flight case2

General

Gas filling pressureCylindrical BF₃ detector, 0.26 atm2
CompatibilitySATURN ratemeter that displays the H*(10) rate and integrated values and compares them with alarm thresholds1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ELSE NUCLEARProduct page Product page Checked 2026-10-16
  2. 2
    Publisher: ELSE NUCLEARDatasheet p. 2, Technical Specifications; Datasheet Checked 2026-10-16
ELSE NUCLEAR SATURN II
ELSE NUCLEAR

ELSE NUCLEAR SATURN II

Digital ratemeter/controller

Format
Wall-mounted GMP version without external cables or connectors
Display
LCD, 2 × 16 characters, backlit
Operating temperature
−25 to +50 °C
11 referenced values
SATURN 5702
ELSE NUCLEAR

SATURN 5702

Mobile gamma and neutron monitor

Weight
Approximately 130 kg
13 referenced values
Neutron Area Monitor
Rotem

Neutron Area Monitor

Wall-mounted neutron detector (Ludlum 42-30H) for area monitoring in WebiSmarts

Detector
Proportional neutron detector Ludlum Model 42-30H; moderator: polyethylene sphere ø25.4 cm
Energy range
Thermal to approx. 12 MeV
Weight
8.8 kg (19.5 lb)
11 referenced values