RDS Base Unit
Survey meter and common base unit for the RDS probes
- Detector type
- Low-dose and high-dose Geiger-Mueller tubes
- Energy range
- Photons 60 keV – 3 MeV; beta 200 keV – 3 MeV
- IP rating
- IP67

Bild: Ludlum
GM pancake probe for contamination monitoring
Small handheld probe for the RDS Base Unit with a GM pancake tube and a 15.5 cm² detection area, for contamination monitoring. It comes with a removable clip-on cover for the front face; an aluminium-clad lead shield for high background areas is available as an option.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-11-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Detector type | Geiger-Mueller pancake tube1 |
|---|---|
| Window | 15.5 cm² mica, protected by a stainless steel screen1 |
| Background rate | 60 cpm with shield, 100 cpm without (at 0.25 µSv/h)1 |
|---|
| Power supply | Supplied by the RDS Base Unit1 |
|---|
| Operating temperature | −30 to 50 °C1 |
|---|---|
| Storage temperature | −50 to 60 °C1 |
| Humidity | 3–100% relative humidity1 |
| IP rating | IP671 |
| Part number | V0388201 |
|---|
| Minimum detectable concentration | 100 Bq/cm²1 |
|---|---|
| Standards met | ANSI N42.17 and N42.34; MIL-STD 461F, 1686C and 810G1 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Survey meter and common base unit for the RDS probes

Alpha/beta probe for surface contamination

Air proportional alpha detector with a 100 cm² sensitive area

Pancake GM probe for alpha, beta and gamma surface contamination
We help you choose the right configuration, detector and accessories for your measurement task.