RDS SABG Probe

GM pancake probe for contamination monitoring

Small handheld probe for the RDS Base Unit with a GM pancake tube and a 15.5 cm² detection area, for contamination monitoring. It comes with a removable clip-on cover for the front face; an aluminium-clad lead shield for high background areas is available as an option.

Detector type
Geiger-Mueller pancake tube1
Energy range
Beta 100 keV – 5 MeV1
Count-rate range
0–1.5 Mcpm1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-11-16

Manufacturer’s product page (opens in a new tab)

Already have the instrument? Request service or support →

Overview

Key features

  • GM pancake tube with 15.5 cm² detection area
  • Removable clip-on cover
  • Optional lead shield for high background

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Detector typeGeiger-Mueller pancake tube1
Window15.5 cm² mica, protected by a stainless steel screen1

Performance

Background rate60 cpm with shield, 100 cpm without (at 0.25 µSv/h)1

Measuring range

Energy rangeBeta 100 keV – 5 MeV1
Count-rate range0–1.5 Mcpm1

Power & battery

Power supplySupplied by the RDS Base Unit1

Mechanical

Dimensions14 × 7 × 12 cm (5.4 × 2.7 × 4.7 in)1
Housing materialAnodised aluminium1

Environment & ratings

Operating temperature−30 to 50 °C1
Storage temperature−50 to 60 °C1
Humidity3–100% relative humidity1
IP ratingIP671

Ordering & accessories

Part numberV0388201

General

Minimum detectable concentration100 Bq/cm²1
Standards metANSI N42.17 and N42.34; MIL-STD 461F, 1686C and 810G1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: Ludlum MeasurementsDatasheet p. 2, Radiological; p. 2, Radiological – Detection; p. 2, Radiological – Background; p. 2, Mechanical … Checked 2026-11-16
RDS Base Unit
Ludlum

RDS Base Unit

Survey meter and common base unit for the RDS probes

Detector type
Low-dose and high-dose Geiger-Mueller tubes
Energy range
Photons 60 keV – 3 MeV; beta 200 keV – 3 MeV
IP rating
IP67
16 referenced values
RDS Alpha-Beta Probe
Ludlum

RDS Alpha-Beta Probe

Alpha/beta probe for surface contamination

Active area
100 cm²
Energy range
Alpha 3–8 MeV
Count-rate range
0–1.5 Mcpm
16 referenced values
PA-100 Detector
Rotem

PA-100 Detector

Air proportional alpha detector with a 100 cm² sensitive area

Detector
Air proportional alpha detector
Weight
1 kg
9 referenced values
GM-10 Detector
Rotem

GM-10 Detector

Pancake GM probe for alpha, beta and gamma surface contamination

Detector
Pancake GM tube LND 73118 or equivalent; mica window 1.5–2.0 mg/cm² behind a fine stainless-steel wire mesh
Count-rate range
0–42,000 cps
Weight
370 g (0.8 lb)
8 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.