RDS Base Unit
Survey meter and common base unit for the RDS probes
- Detector type
- Low-dose and high-dose Geiger-Mueller tubes
- Energy range
- Photons 60 keV – 3 MeV; beta 200 keV – 3 MeV
- IP rating
- IP67

Bild: Ludlum
Alpha/beta probe for surface contamination
Probe for the RDS Base Unit that measures alpha and beta surface contamination simultaneously, with a ZnS(Ag) scintillator on a thick plastic scintillator. It carries its own calibration and has a removable protective grate that can be decontaminated.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-11-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Background rate | In a 10 µR/h field: alpha < 3.0 cpm, beta < 300 cpm1 |
|---|
| Energy range | Alpha 3–8 MeV1 |
|---|---|
| Count-rate range | 0–1.5 Mcpm1 |
| Gamma sensitivity | > 35 cps per µGy/h1 |
| Power supply | Supplied by the RDS Base Unit1 |
|---|
| Operating temperature | −30 to 50 °C1 |
|---|---|
| Storage temperature | −50 to 60 °C1 |
| Humidity | 3–100% relative humidity1 |
| IP rating | IP54 (a contaminated Mylar window should be replaced)1 |
| Cross talk | Alpha to beta (Am-241) < 15%; beta to alpha (Sr-90/Y-90) < 1%1 |
|---|---|
| Standards met | ANSI N42.17 and N42.34; MIL-STD 461F, 1686C and 810G1 |
| NATO Stock Number (NSN) | 6665-01-671-44791 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Survey meter and common base unit for the RDS probes

GM pancake probe for contamination monitoring

Air proportional alpha detector with a 100 cm² sensitive area

Pancake GM probe for alpha, beta and gamma surface contamination