PARSTAT MC

Modular multichannel system

Configure multiple electrochemical measurement channels for testing and research. PARSTAT MC combines replaceable modules with VersaStudio control.

Measurement channels
Up to 20 with PMC-200; up to 10 with PMC-1000 or PMC-2000A1
Compliance voltage
PMC-200: ±10 V; PMC-1000: ±12 V; PMC-2000A: ±30 V1

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Chassis supports up to 20 channels.
  • Modules address high current, high voltage or two channels per card.
  • EIS up to 7 MHz with PMC-2000A.

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Measuring range

Polarization voltagePMC-200 / PMC-1000: ±10 V; PMC-2000A: ±30 V1
EIS frequency rangePMC-200: 1 mHz–100 kHz; PMC-1000: 10 µHz–1 MHz; PMC-2000A: 10 µHz–7 MHz1

Electronics & signal

Measurement channelsUp to 20 with PMC-200; up to 10 with PMC-1000 or PMC-2000A1
Maximum output currentPMC-200: ±1 A per channel; PMC-1000: ±2 A; PMC-2000A: ±1 A1

Software & data

SoftwareVersaStudio included1

Connectivity & interface

Cell connectionsCE, WE, RE, SE; PMC-2000A has 6-wire capability with 1 auxiliary input1

Environment & ratings

Compliance voltagePMC-200: ±10 V; PMC-1000: ±12 V; PMC-2000A: ±30 V1

Ordering & accessories

External current boosterPMC-1000 / PMC-2000A: optional ±10 / ±20 / ±30 A1

General

Chassis slots10, USB chassis1
Floating measurement / isolationFloating, isolated on PMC-200, PMC-1000 and PMC-2000A1
Host systemPARSTAT MC USB chassis, 10 slots1
Floating measurementIsolated and floating1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Models

3 models in the series. Filter or scroll the table to compare.

3 / 3
ModelMeasurement channels per card1Maximum current without boosterPolarization voltageCompliance voltage2Best voltage resolution1Best current resolution without LCI1Maximum acquisition rateEIS frequency range2LCI resolution with option1Auxiliary channels1
PMC-2002±1 A per channel2±10 V2±10 V24-bit; 1.2 µV24-bit; 240 fA250 kS/s21 mHz–100 kHz––
PMC-10001±2 A2±10 V2±12 V16-bit; 6 µV16-bit; 122 fA1 MS/s110 µHz–1 MHz122 aA with VersaSTAT-LC–
PMC-2000A1±1 A1±30 V1±30 V18-bit; 916 nV18-bit; 122 fA1 MS/s110 µHz–7 MHz–1 differential channel, DC/EIS; 6-wire

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: AMETEK Scientific InstrumentsProduct page Differentiated Specifications; product page; Differentiated Specifications – Grounding; Differentiated Specifications – PMC-2000A column Checked 2026-10-16
  2. 2
    Publisher: AMETEK Scientific InstrumentsBrochure p. 2; p. 3; p. 6 Checked 2026-10-16
ModuLab XM MTS materialtestsystem
AMETEK SI

ModuLab XM MTS materialtestsystem

Materials test system

Impedance range
10 µΩ–100 TΩ depending on module configuration
EIS frequency range
10 µHz–1 MHz with XM MFRA 1MHz
Maximum output current
XM MAT: ±100 mA; XM MBST 2A: ±2 A (option)
6 models
ModuLab XM ECS
AMETEK SI

ModuLab XM ECS

Configurable electrochemical system

Compliance voltage
±8 V standard; ±30 V / ±100 V with HV options
8 models
XM CHAS 08
AMETEK SI · Solartron Analytical

XM CHAS 08

System chassis

Slots
8
Compatible systems
ModuLab XM ECS / MTS
Module groups
ECS and MTS groups can run independently in the same chassis
3 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.