ORTEC 567

Time-to-amplitude converter / SCA

Converts the interval between start and stop pulses to an analog amplitude for time spectroscopy. An integrated SCA selects time intervals.

Base full-scale ranges
50, 100, 200 ns1
Range multipliers
1, 10, 100, 1000, 10 0001
Maximum full-scale range
2 ms1
Time resolution (FWHM)
≤0.01% of full scale + 5 ps, all ranges1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Selectable time ranges
  • Internal or external strobe
  • Valid Start and Valid Conversion outputs

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Performance

Time resolution (FWHM)≤0.01% of full scale + 5 ps, all ranges1
Integral nonlinearity≤±0.1%, from 10 ns or 2% full scale (whichever is greater) to 100% full scale1
Typical differential nonlinearity<1%, from 10 ns or 2% full scale (whichever is greater) to 100% full scale1
SCA threshold nonlinearity≤±0.5% of full scale1

Measuring range

Base full-scale ranges50, 100, 200 ns1
Range multipliers1, 10, 100, 1000, 10 0001
Maximum full-scale range2 ms1

Electronics & signal

Output delay, internal strobe0.5–10.5 µs1
TAC output amplitude0…+10 V1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ORTEC (AMETEK)Datasheet p. 1, Controls (RANGE); p. 1, Controls (MULTIPLIER); p. 1, features/description; p. 1, Specifications – Performance … Checked 2026-10-16
ORTEC 276L
ORTEC

ORTEC 276L

Low-power PMT base with preamplifier

PMT compatibility
10-stage, 14-pin; TRW 3B14 socket (JEDEC B14-38); directly compatible with many integrated NaI-PMT assemblies
Conversion gain
Nominally 5 µV/eV with a 2 × 2 in NaI(Tl) crystal and PMT gain of 10⁶
Weight
0.65 kg net
14 referenced values
ORTEC 269
ORTEC

ORTEC 269

PMT base for 14- and 12-stage, 20-pin photomultipliers

PMT compatibility
14- or 12-stage, 20-pin socket (JEDEC B20-102), including Hamamatsu R1250; Philips XP2020, XP2020Q, XP2040, XP2040Q, XP2041, XP2041Q, XP2212B, XP2230B, XP2233B, XP2262B; EMI 9813K
Weight
0.63 kg net
6 referenced values
ORTEC 266
ORTEC

ORTEC 266

PMT base for 10-stage, 14-pin photomultipliers

PMT compatibility
10-stage, 14-pin; TRW 3B14 socket (JEDEC B14-38)
Weight
0.37 kg net
6 referenced values
ORTEC 265A
ORTEC

ORTEC 265A

PMT base for 12-stage, 21-pin photomultipliers, for fast timing

PMT compatibility
12-stage, standard 21-pin sockets, including Hamamatsu R329, R1332, R1333 and Burle 8575, 8850, C31000M
Weight
0.63 kg net
6 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.