ORTEC 276
PMT base with preamplifier
- PMT compatibility
- 10-stage, 14-pin; JEDEC B14-38
- Weight
- 0.65 kg net

Bild: ORTEC
Low-power PMT base with preamplifier
PMT base with integral low-noise preamplifier, low-power voltage divider and focus control for scintillation detectors, ideally NaI(Tl). The preamplifier output is used for energy analysis and the anode output for timing or auxiliary energy analysis; a test input accepts pulses from a pulse generator.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-17
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| PMT compatibility | 10-stage, 14-pin; TRW 3B14 socket (JEDEC B14-38); directly compatible with many integrated NaI-PMT assemblies1 |
|---|
| Integral nonlinearity | <±0.02% from 0 to +10 V1 |
|---|---|
| Output noise | <50 µV rms with an ORTEC main amplifier (e.g. Model 672) and 1 µs time constant1 |
| Preamplifier | Integral low-noise preamplifier, dc-coupled positive output (Z₀ = 93 Ω) for energy analysis1 |
|---|---|
| Voltage divider | Total 5.6 MΩ, giving a bleeder current of 200 µA at a typical 1 kV; linear distribution to all stages1 |
| Detector bias | External positive high voltage: +2,000 V maximum (SHV); observe the rated voltage of the installed tube1 |
| Output rise time | <100 ns for test input or fast scintillator1 |
| Conversion gain | Nominally 5 µV/eV with a 2 × 2 in NaI(Tl) crystal and PMT gain of 10⁶1 |
| Signal outputs | Preamp: positive pulse, Z₀ = 93 Ω, dc-coupled (BNC); anode: negative pulse for timing or auxiliary energy analysis, Z₀ = 1 kΩ, ac-coupled (BNC)1 |
| Power supply | Preamplifier: +24 V, 16 mA and −24 V, 16 mA via a captive 4 m cable (Amphenol 17-20090); compatible with all ORTEC main amplifiers and the 4002P1 |
|---|
| Temperature instability | <±0.005%/°C, 0 to 50 °C1 |
|---|
| Decay time | 50 µs time constant1 |
|---|
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

PMT base with preamplifier


PMT base for 14- and 12-stage, 20-pin photomultipliers

PMT base for 10-stage, 14-pin photomultipliers
We help you choose the right configuration, detector and accessories for your measurement task.