NAUSICAA 2IC
Gaseous effluent monitor
- Detector type
- 2 identical ionization chambers
- Total weight
- Approximately 150 kg

Bild: ELSE NUCLEAR
Pressurised ion chamber gamma monitor, 10 nSv/h–10 Sv/h
Gamma monitoring unit with a pressurised ion chamber, electrometer and CPU-based control unit, with very high sensitivity and stable response. IC-T is intended for environmental measurements and IC-T-PF for pulsed fields around accelerators. ICP-T is the same detector managed by a SATURN ratemeter, and DISCOVERY IC-T is an IP65 version for outdoor use.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-17
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Detector type | Pressurised ion chamber, argon 8 atm + xenon 2 atm, 10 L total volume1 |
|---|
| Accuracy | ±5% at environmental background radiation levels1 |
|---|
| Dose-rate range | 10 nSv/h–10 Sv/h (standard); 10 nSv/h–100 mSv/h (PF pulsed-field version)1 |
|---|---|
| Energy range | 30 keV–10 MeV1 |
| Ion chamber sensitivity | Approx. 2 × 10⁻⁸ A/(R/h)1 |
| Communication | Ethernet or serial connection to a host PC running 5700 sMON2 |
|---|
| Weight | 21 kg (NAUSICAA IC-T); 16 kg (ICP-T)1 |
|---|---|
| Dimensions | 365 × 390 × 600 mm (NAUSICAA IC-T); 256 × 327 × 600 mm (ICP-T) (W × D × H)1 |
| Operating temperature | −25 to +50 °C1 |
|---|
| Configurations | NAUSICAA IC-T with CPU-based control unit; ICP-T without CPU, managed by a SATURN ratemeter; DISCOVERY IC-T in an IP65 housing for outdoor use with wireless data transfer2 |
|---|---|
| Accessories | Traceable calibration (¹³⁷Cs), E-N-KEY external keyboard and ALU alarm unit (IC-T), trolley, wall mounting kit, IP54 enclosure, GPS and photovoltaic panel (DISCOVERY IC-T)1 |
| Electronics | Two alternative electrometers with automatic scale change: 9 decades for environmental measurements (IC-T), 7 decades for pulsed fields around accelerators (IC-T-PF)1 |
|---|
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Gaseous effluent monitor