ICS Integrated Cryocooling System

Mechanical cooling for HPGe

Cool HPGe detectors without liquid nitrogen. ICS uses a Stirling cooler and is available in configurations for laboratory and field operation.

Weight
E/HJ: 14 kg excluding detector; P4: 12.7 kg excluding detector capsule1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Compatible with ORTEC GEM, GMX and PROFILE detectors.
  • Three versions with different cryostat configurations.
  • Active vibration cancellation and maintenance-free mechanical cooling.

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Detector compatibilityORTEC GEM, GMX and PROFILE1

Performance

Audible noise<55 dBA at 1 m and ambient <30 °C1
Effect on energy resolutionAbove 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂1

Cooling

Cooling technologySunpower Stirling cooler with active vibration cancellation (AVC)1

Power & battery

Power supply100–240 V AC, 50/60 Hz1
Power consumptionE/HJ: 70 W typical, 130 W max; P4: 80 W typical, 200 W max at room temperature1
Backup batteryNone built-in; external battery power or UPS may be used1

Mechanical

WeightE/HJ: 14 kg excluding detector; P4: 12.7 kg excluding detector capsule1

Environment & ratings

Operating temperature−10 to +40 °C1

Ordering & accessories

ModelsICS-E, ICS-HJ and ICS-P41

General

PopTop compatibilityICS-P4: yes; ICS-E/HJ: no1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Models

4 models in the series. Filter or scroll the table to compare.

4 / 4
ModelCooling technology2Detector compatibility2Power supply2Power consumption2Operating temperature2Humidity2Audible noise2PopTop compatibility2Hardened cryostat2Compatible endcap windows2Molecular sieve2Operating orientation2Effect on energy resolution2Status indication2Detector connectors2Stand2Ordering code2Height adjustment2Compatibility2Installation into shielding2Stand weight3Maximum combined load3Periodic stand maintenance3Weight2Cooling rod length (selected when ordering)2Ordering4Preamplifier option4
CFG-ICS-P4Sunpower Stirling cooler with active vibration cancellation (AVC)ORTEC GEM, GMX and PROFILE; final combination selected when ordering100–240 V AC, 50/60 Hz80 W typical; 200 W maximum at room temperature−10 to +40 °CNon-condensing<55 dBA at 1 m, fully operational and ambient <30 °CYesNo; conventional cryostatAluminium or carbon fibreYesAll attitudesAbove 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂1 LED: power onNone; detector cable set not includedICS STAND 2 is optional and not includedCFG-ICS-P4; cooler version ICS-P4––––––––––
ICS STAND 2–––––––––––Vertical only–––––Multiple height positions with pneumatic arm for raising/loweringICS-E and ICS-P4 explicitly listed in ordering table; brochure narrative states any ICS modelInstalled from below; opening must exceed ICS-E detector seal-clamp diameter26 lb (approx. 11.8 kg)100 lb including ICS, detector and any lead plugsNot required; detector and cooler have their own maintenance requirements––––
ICS-ESunpower Stirling cooler with active vibration cancellation (AVC)ORTEC GEM, GMX and PROFILE; final combination selected when ordering100–240 V AC, 50/60 Hz70 W typical; 130 W maximum at room temperature−10 to +40 °CNon-condensing<55 dBA at 1 m, fully operational and ambient <30 °CNoYes; can recool without full thermal cycling after partial warm-upAluminiumNoAll attitudesAbove 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂2 LEDs: power on, cooler on0.3 m detector cable bundle from detector endcap; detector cable set includedICS STAND 2 is optional and not included–––––––14 kg excluding detector25, 114, 139 or 190 mm (tolerance ±3 mm); other lengths by special orderStreamline only: suffix -ICS-E on the detector model, plus CFG-ICS-E and a cooling-rod line item ICS-LLExxx-[endcap]; sold only with a detectorPLUS ultra-high-count-rate preamplifier available as CFG-ICS-E-PL (ICS with external PLUS preamp); cooling rod length must be specified
ICS-HJSunpower Stirling cooler with active vibration cancellation (AVC)ORTEC GEM, GMX and PROFILE; final combination selected when ordering100–240 V AC, 50/60 Hz70 W typical; 130 W maximum at room temperature−10 to +40 °CNon-condensing<55 dBA at 1 m, fully operational and ambient <30 °CNoYes; can recool without full thermal cycling after partial warm-upAluminiumNoAll attitudesAbove 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂3 LEDs: power on, cooler on, status on preamp6 BNC, 1 SHV and 1 nine-pin connector on preamp; detector cable set includedICS STAND 2 is optional and not included–––––––14 kg excluding detector–Integrated with selected detector using suffix -ICS-HJ; not a standalone PopTop cooler–

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ORTEC (AMETEK)Product page Specifications – Common; More Information; Specifications – Unique; Specifications – Common (Environmental) … Checked 2026-10-16
  2. 2
    Publisher: ORTEC (AMETEK)Brochure p. 5 'Specifications – Common for ICS-E, ICS-HJ, and ICS-P4' (Cooler); p. 2 Introduction; p. 5 'Specifications – Common…' (HPGe Compatibility); p. 7 'ICS-P4 GEM, GMX, and PROFILE PopTop Detector Dimensions'; p. 5 'Specifications – Common…' (Electrical Supply); p. 6 'Specifications – Unique…', ICS-P4 column … Checked 2026-10-16
  3. 3
    Publisher: ORTEC (AMETEK)Manual PDF p. 6 (manual p. 4), §1.1.2 Mechanical; PDF p. 6 (manual p. 4), §1.1.1 Performance Checked 2026-10-16
  4. 4
    Publisher: ORTEC (AMETEK)Product catalogue PDF p. 13 'GEM Detector Options' (-ICS-E) and PDF p. 15 'GEM Streamline ICS Cryostat'; PDF p. 15 'GEM Streamline ICS Cryostat' (also PDF p. 18 for PROFILE); PDF p. 13 'GEM Detector Options' (-ICS-HJ); PDF p. 17 for PROFILE Checked 2026-10-16
Möbius Recycler
ORTEC

Möbius Recycler

HPGe detector cooling

Weight
28.58 kg empty; 51.26 kg full, excluding cryostat and detector
LN₂ hold time during power loss
Full dewar, typical detector: Streamline about 7–10 days; PopTop about 4–7 days
4 models
Shallow Hole Probe
ORTEC

Shallow Hole Probe

HPGe gamma detector

Detector type
Configured with GEM, GMX, GLP, PROFILE or Safeguards HPGe
Weight
Approximately 12 lb
LN₂ holding time
Approximately 16 hours with 70 mm endcap
1 model
Dewars and Cryostats
ORTEC

Dewars and Cryostats

HPGe detector cooling

Cooling technology
Liquid nitrogen (LN₂)
Cryostat configurations
Vertical and horizontal dipstick, portable (all-position or multi-orientation Gamma Gage), down-looking, side-looking and J-type, depending on detector configuration
Dewar capacity
Gamma Gage: 1.2/3/5 L; MOD multi-orientation: 3/7 L; side-looking and down-looking: 7.5/13/30 L; dipstick cryostats: 30 L (DWR-30); shallow-hole probe: 0.7 L
43 models
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.