Möbius Recycler
HPGe detector cooling
- Weight
- 28.58 kg empty; 51.26 kg full, excluding cryostat and detector
- LN₂ hold time during power loss
- Full dewar, typical detector: Streamline about 7–10 days; PopTop about 4–7 days

Bild: ORTEC
Mechanical cooling for HPGe
Cool HPGe detectors without liquid nitrogen. ICS uses a Stirling cooler and is available in configurations for laboratory and field operation.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Detector compatibility | ORTEC GEM, GMX and PROFILE1 |
|---|
| Audible noise | <55 dBA at 1 m and ambient <30 °C1 |
|---|---|
| Effect on energy resolution | Above 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂1 |
| Cooling technology | Sunpower Stirling cooler with active vibration cancellation (AVC)1 |
|---|
| Power supply | 100–240 V AC, 50/60 Hz1 |
|---|---|
| Power consumption | E/HJ: 70 W typical, 130 W max; P4: 80 W typical, 200 W max at room temperature1 |
| Backup battery | None built-in; external battery power or UPS may be used1 |
| Weight | E/HJ: 14 kg excluding detector; P4: 12.7 kg excluding detector capsule1 |
|---|
| Operating temperature | −10 to +40 °C1 |
|---|
| Models | ICS-E, ICS-HJ and ICS-P41 |
|---|
| PopTop compatibility | ICS-P4: yes; ICS-E/HJ: no1 |
|---|
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
4 models in the series. Filter or scroll the table to compare.
| Model | Cooling technology2 | Detector compatibility2 | Power supply2 | Power consumption2 | Operating temperature2 | Humidity2 | Audible noise2 | PopTop compatibility2 | Hardened cryostat2 | Compatible endcap windows2 | Molecular sieve2 | Operating orientation2 | Effect on energy resolution2 | Status indication2 | Detector connectors2 | Stand2 | Ordering code2 | Height adjustment2 | Compatibility2 | Installation into shielding2 | Stand weight3 | Maximum combined load3 | Periodic stand maintenance3 | Weight2 | Cooling rod length (selected when ordering)2 | Ordering4 | Preamplifier option4 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CFG-ICS-P4 | Sunpower Stirling cooler with active vibration cancellation (AVC) | ORTEC GEM, GMX and PROFILE; final combination selected when ordering | 100–240 V AC, 50/60 Hz | 80 W typical; 200 W maximum at room temperature | −10 to +40 °C | Non-condensing | <55 dBA at 1 m, fully operational and ambient <30 °C | Yes | No; conventional cryostat | Aluminium or carbon fibre | Yes | All attitudes | Above 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂ | 1 LED: power on | None; detector cable set not included | ICS STAND 2 is optional and not included | CFG-ICS-P4; cooler version ICS-P4 | – | – | – | – | – | – | – | – | – | – |
| ICS STAND 2 | – | – | – | – | – | – | – | – | – | – | – | Vertical only | – | – | – | – | – | Multiple height positions with pneumatic arm for raising/lowering | ICS-E and ICS-P4 explicitly listed in ordering table; brochure narrative states any ICS model | Installed from below; opening must exceed ICS-E detector seal-clamp diameter | 26 lb (approx. 11.8 kg) | 100 lb including ICS, detector and any lead plugs | Not required; detector and cooler have their own maintenance requirements | – | – | – | – |
| ICS-E | Sunpower Stirling cooler with active vibration cancellation (AVC) | ORTEC GEM, GMX and PROFILE; final combination selected when ordering | 100–240 V AC, 50/60 Hz | 70 W typical; 130 W maximum at room temperature | −10 to +40 °C | Non-condensing | <55 dBA at 1 m, fully operational and ambient <30 °C | No | Yes; can recool without full thermal cycling after partial warm-up | Aluminium | No | All attitudes | Above 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂ | 2 LEDs: power on, cooler on | 0.3 m detector cable bundle from detector endcap; detector cable set included | ICS STAND 2 is optional and not included | – | – | – | – | – | – | – | 14 kg excluding detector | 25, 114, 139 or 190 mm (tolerance ±3 mm); other lengths by special order | Streamline only: suffix -ICS-E on the detector model, plus CFG-ICS-E and a cooling-rod line item ICS-LLExxx-[endcap]; sold only with a detector | PLUS ultra-high-count-rate preamplifier available as CFG-ICS-E-PL (ICS with external PLUS preamp); cooling rod length must be specified |
| ICS-HJ | Sunpower Stirling cooler with active vibration cancellation (AVC) | ORTEC GEM, GMX and PROFILE; final combination selected when ordering | 100–240 V AC, 50/60 Hz | 70 W typical; 130 W maximum at room temperature | −10 to +40 °C | Non-condensing | <55 dBA at 1 m, fully operational and ambient <30 °C | No | Yes; can recool without full thermal cycling after partial warm-up | Aluminium | No | All attitudes | Above 100 keV: warranted FWHM unaffected; below 100 keV: FWHM limit +10% versus LN₂ | 3 LEDs: power on, cooler on, status on preamp | 6 BNC, 1 SHV and 1 nine-pin connector on preamp; detector cable set included | ICS STAND 2 is optional and not included | – | – | – | – | – | – | – | 14 kg excluding detector | – | Integrated with selected detector using suffix -ICS-HJ; not a standalone PopTop cooler | – |
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

HPGe detector cooling

HPGe gamma detector

HPGe detector cooling
We help you choose the right configuration, detector and accessories for your measurement task.