bPAD+VR

bPAD+ with an analogue output for pulses or count rate, e.g. for flow instruments

Enhanced version of the bPAD+ in which the analogue output can deliver either the pulse signal or a voltage proportional to the discriminated count rate. This is useful, for example, in flow count-rate instruments and HPLC systems.

Operating modes
Differential and integral1
Discriminator
Upper and lower level discriminator, 2,048 steps1
Gain
×1–×16 (×1–×256 on request)1
Signal outputs
TTL-compatible pulse output of programmable duration1

Specifications checked against the manufacturer’s documents on 2026-10-17

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Differential or integral mode
  • Programmable response time of the rate output
  • External 9–36 V supply

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Operating modesDifferential and integral1
DiscriminatorUpper and lower level discriminator, 2,048 steps1
Gain×1–×16 (×1–×256 on request)1
Signal outputsTTL-compatible pulse output of programmable duration1
Analog outputPulse signal 0–2 V or a voltage of 0–2.5 V proportional to the discriminated count rate (50 Ω, variable sensitivity); response time 0.1–20 s in 14 steps1
Detector high voltageUp to 1,500 V (photomultiplier)1
Supply voltageExternal 9–36 V1
Dimensionsø56 × 60 mm (71 mm including the SMA connectors)1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: BrightSpecProduct page Product page – Main features; Product page Checked 2026-10-16
ILLUSTRATION
BrightSpec

bPAD+

bPAD with an extra analogue pulse output to an external MCA

Operating modes
Differential and integral
Discriminator
Upper and lower level discriminator, 2,048 steps
Gain
×1–×256
8 referenced values
bPAD
BrightSpec

bPAD

Digital single-channel analyser (SCA) in the tube socket, controlled via USB

Weight
Approx. 90 g
9 referenced values
bDAS
BrightSpec

bDAS

Digital acquisition system for synchronised counting with up to 12 probes

Weight
Probe: approx. 4 kg
6 referenced values
bPAD 422
BrightSpec

bPAD 422

Digital single-channel analyser in the tube socket with RS422/485, for radiotracer measurements

Weight
Approx. 80 g
8 referenced values
Contact us

Questions about this product?