FOOMON
Food monitoring
- Detector type
- 2 × 2 inch NaI(Tl)
- Total weight
- Approximately 22 kg
- IP rating
- IP67 during transport with closed lid; IP65 in operating setup

Bild: ELSE NUCLEAR
Emergency screening
Portable thyroid monitor for emergency use – fast, reliable measurement of I-131 in the thyroid with a collimated NaI(Tl) detector. Compliant with IEC 61582.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Signal processing | SiPM matrix and compact MCA1 |
|---|
| Detector probe weight | 3.5 kg1 |
|---|---|
| Total weight | Approximately 19 kg1 |
| Transport case dimensions | 62.8 × 49.2 × 22.3 cm1 |
| Operating temperature | −20 to +50 °C1 |
|---|---|
| IP rating | IP67 during transport with closed lid; IP65 in operating setup1 |
| Lead collimator | 15 mm1 |
|---|---|
| MDA, typical in 2-minute screening | 90–120 Bq1 |
| Maximum measurable activity | Over 3 MBq1 |
| Standard | Compliant with IEC 615821 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Food monitoring

Surface contamination monitor for pipe sections

Conveyor-belt contamination monitor for small objects

Contamination monitor for bags with removable detectors
We help you choose the right configuration, detector and accessories for your measurement task.