Ludlum Model 733
PC-based gamma MCA system
- Part number
- 48-3835
- NaI(Tl) crystal (Ø × L)
- 7.6 × 7.6 cm
- Spectrum channels
- 1024

Bild: Ludlum
PC-based gamma MCA system
USB-powered gamma spectroscopy system with a NaI(Tl) detector, MCA and SAAS software. Computer not included.
Download product sheet (PDF)All values with references – for tender documents
Specifications checked against the manufacturer’s documents on 2026-10-16
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.
| Spectrum channels | 10241 |
|---|---|
| ADC | 10-bit Wilkinson1 |
| ADC clock | 48 MHz1 |
| ADC conversion at full scale | 20 µs1 |
| Internal detector high voltage | 0–1200 V DC; 0.5 mA; ripple <30 mV1 |
| Pulse shaping | 1 µs, semi-Gaussian, internal pole-zero adjustment1 |
| Communication | USB; ANSI N42 export1 |
|---|
| USB power | Approximately 4.75 V at 450 mA1 |
|---|
| NaI(Tl) crystal (Ø × L) | 5.1 × 5.1 cm1 |
|---|---|
| USB cable length | 5 m1 |
| Dimensions | Detector: 6.4 × 20.3 cm (D × L)2 |
| Weight | 0.9 kg2 |
The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.
The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.