Ludlum Model 3003

SCA and detector readout

Readout instrument for external radiation detectors with a single-channel analyzer and timed counting. Base model 3003 has no internal detector.

Weight excluding detectors and clips
0.9 kg1
Enclosure rating and conditions
IP531
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Key features

  • Selectable RATE, MAX, COUNT and DOSE modes.
  • Separate setups for up to four external detectors.
  • Logging requires the optional Lumic Data Logger Kit.

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

External detector setupsUp to 4; GM, scintillation, proportional or neutron1
Compatible external detectorsUp to 4 external GM, scintillation, proportional or neutron detectors1

Performance

Linearity±10 %1

Measuring range

Response timeUser-selectable from 1 to 60 seconds, auto SLOW, auto FAST, fixed SLOW (22 s), fixed FAST (4 s)1

Electronics & signal

Detector high voltage300–2400 V DC1
Detector bias300 to 2400 Vdc1

Connectivity & interface

Count display range0–999 million counts; operational range depends on detector and dead time1
CommunicationUSB; optional Bluetooth/RS-232/TTL, select one output option2

Power & battery

Approximate battery life175 h1
Power supply4 AA batteries1
Battery lifeApproximately 175 hours (series specification)1

Mechanical

Dimensions (H × W × L)16.5 × 11.4 × 21.6 cm1
Weight excluding detectors and clips0.9 kg1

Environment & ratings

Operating temperature−20–+50 °C1
Enclosure rating and conditionsIP531

Ordering & accessories

Part number48-42891
Logging capacity (optional)Up to 1000 data points; Lumic Data Logger Kit required1

General

AnalyzerSCA with upper window2

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Models

5 models in the series. Filter or scroll the table to compare.

5 / 5
ModelPart number1Internal detector1Internal detector range1Dose response condition1
3003i48-4326Energy compensated GM1 µSv/h to 50 mSv/h (0.1 mR/hr to 5 R/hr)–
3003i/148-4378CsI scintillator, 18 x 18 mmBackground to 500 µSv/h (50 mR/hr)Nonlinear energy response; dose and exposure readings may be over- or under-reported
3003i/348-4411Energy compensated GM1 µSv/h to 250 mSv/h (0.1 mR/hr to 25 R/hr)–
3003i/448-4326-4Energy compensated GM10 µSv/h to 500 mSv/h (1 mR/hr to 50 R/hr)–
3003i/648-4326-6Energy compensated GM40 µSv/h to 10 Sv/h (4 mR/hr to 1000 R/hr)–

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: Ludlum MeasurementsDatasheet p. 1, Configurations table; p. 2, Specifications (COMPATIBLE DETECTORS); p. 2, Specifications; p. 2, Specifications (DISPLAY RANGE and NOTE) … Checked 2026-10-16
  2. 2
    Publisher: Ludlum MeasurementsProduct page Introduction; Features; Options tab (Data Output Options) Checked 2026-10-16
Ludlum Model 2360
Ludlum

Ludlum Model 2360

Alpha/beta scaler with data logging

Weight including batteries
1.6 kg
11 referenced values
Ludlum Model 2350-1
Ludlum

Ludlum Model 2350-1

Scaler/ratemeter with data logging

Weight including batteries
2.4 kg
12 referenced values
Ludlum Model 2241-3
Ludlum

Ludlum Model 2241-3

Digital scaler/ratemeter

Weight including batteries
1.6 kg
12 referenced values
Ludlum Model 2241-2
Ludlum

Ludlum Model 2241-2

Digital scaler/ratemeter

Weight including batteries
1.6 kg
12 referenced values