Alpha Suite alfaspektrometrar

Integrated alpha spectrometry

Alpha Suite alfaspektrometrar – integrated alpha spectrometry from ORTEC.

Energy resolution
≤20 keV FWHM with BU-017-450-100 ULTRA detector and ²⁴¹Am point source; detector-to-source spacing equal to detector diameter1
Spectrum channels
Software-selectable conversion gain: 256, 512, 1024, 2048 or 4096 channels1
Datasheet

Specifications checked against the manufacturer’s documents on 2026-10-16

Manufacturer’s product page (opens in a new tab)

Overview

Specifications

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

Detector

Maximum detector areaAria/Duo: 1200 mm²; Mega: 3000 mm² (consult factory)1
Maximum sample-to-detector spacing44 mm, in 4 mm increments1

Performance

Energy resolution≤20 keV FWHM with BU-017-450-100 ULTRA detector and ²⁴¹Am point source; detector-to-source spacing equal to detector diameter1

Electronics & signal

Measurement channelsAria: 1; Duo: 2; Mega: 1; Ensemble: up to 4 Duo/Mega modules (maximum 8 channels with Duo)1
Detector bias0 to ±100 V, 10 µA; computer-controlled 1 V increments1
Spectrum channelsSoftware-selectable conversion gain: 256, 512, 1024, 2048 or 4096 channels1

Software & data

SoftwareMAESTRO included1

Connectivity & interface

Vacuum connectionAria/Duo/Mega: 0.25-inch Swagelok; Ensemble: NW251
CommunicationHigh-speed USB, type B1
Detector connectorRear Microdot (ORTEC B mount)1

Mechanical

Maximum sample diameterAria/Duo: 51 mm; Mega: 106 mm1
Chamber materialNickel-plated cast brass1

Environment & ratings

Operating temperature0 to 50 °C1

The data comes from the manufacturer’s published documents. Manufacturers may change specifications; the current configuration and performance are confirmed in the quotation.

Models

19 models in the series. Filter or scroll the table to compare.

19 / 19
ModelMeasurement channels2ConfigurationMaximum sample diameter1Maximum detector area1Recoil protectionVacuum control1Computer connection1Software1Net weight1Host system2Module slots1
ALPHA-ARIA1Single alpha spectroscopy channel in a double-wide NIM chassis151 mm1200 mm²External RCAP-2 required for recoil protection1Manual Pump/Hold/VentUSB type BMAESTRO included; AlphaVision is separate software1.9 kg––
ALPHA-DUO2Benchtop unit with two alpha spectroscopy channels151 mm1200 mm²Built-in pressure controller; optional biased sample trays required1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software7.5 kg––
ALPHA-DUO-M12Dual-input Alpha Ensemble expansion module (two chambers)251 mm1200 mm²Built-in pressure controller; optional biased sample trays required1Computer control per module–––Expansion module for Alpha Ensemble; not a standalone benchtop instrument–
ALPHA-ENSEMBLE-1D-1M3One ALPHA-DUO-M1 module and one ALPHA-MEGA-M1 module2ALPHA-DUO-M1: 51 mm; ALPHA-MEGA-M1: 106 mmALPHA-DUO-M1: 1200 mm²; ALPHA-MEGA-M1: 3000 mm² (consult factory)Built-in RCAP pressure controller for the ALPHA-DUO-M1 module(s) only (requires optional biased sample holders); not available for the ALPHA-MEGA-M1 module(s)1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-1D-2M4One ALPHA-DUO-M1 module and two ALPHA-MEGA-M1 modules2ALPHA-DUO-M1: 51 mm; ALPHA-MEGA-M1: 106 mmALPHA-DUO-M1: 1200 mm²; ALPHA-MEGA-M1: 3000 mm² (consult factory)Built-in RCAP pressure controller for the ALPHA-DUO-M1 module(s) only (requires optional biased sample holders); not available for the ALPHA-MEGA-M1 module(s)1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-1D-3M5One ALPHA-DUO-M1 module and three ALPHA-MEGA-M1 modules2ALPHA-DUO-M1: 51 mm; ALPHA-MEGA-M1: 106 mmALPHA-DUO-M1: 1200 mm²; ALPHA-MEGA-M1: 3000 mm² (consult factory)Built-in RCAP pressure controller for the ALPHA-DUO-M1 module(s) only (requires optional biased sample holders); not available for the ALPHA-MEGA-M1 module(s)1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-1M1One ALPHA-MEGA-M1 module2106 mm3000 mm² (consult factory)RCAP recoil-protection option is for ALPHA-DUO-M1 modules only; biased sample tray not available for Mega modules3Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-22One ALPHA-DUO-M1 module251 mm1200 mm²Built-in pressure controller; optional biased sample trays required1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-2D-1M5Two ALPHA-DUO-M1 modules and one ALPHA-MEGA-M1 module2ALPHA-DUO-M1: 51 mm; ALPHA-MEGA-M1: 106 mmALPHA-DUO-M1: 1200 mm²; ALPHA-MEGA-M1: 3000 mm² (consult factory)Built-in RCAP pressure controller for the ALPHA-DUO-M1 module(s) only (requires optional biased sample holders); not available for the ALPHA-MEGA-M1 module(s)1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-2D-2M6Two ALPHA-DUO-M1 modules and two ALPHA-MEGA-M1 modules2ALPHA-DUO-M1: 51 mm; ALPHA-MEGA-M1: 106 mmALPHA-DUO-M1: 1200 mm²; ALPHA-MEGA-M1: 3000 mm² (consult factory)Built-in RCAP pressure controller for the ALPHA-DUO-M1 module(s) only (requires optional biased sample holders); not available for the ALPHA-MEGA-M1 module(s)1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-2M2Two ALPHA-MEGA-M1 modules2106 mm3000 mm² (consult factory)RCAP recoil-protection option is for ALPHA-DUO-M1 modules only; biased sample tray not available for Mega modules3Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-3D-1M7Three ALPHA-DUO-M1 modules and one ALPHA-MEGA-M1 module2ALPHA-DUO-M1: 51 mm; ALPHA-MEGA-M1: 106 mmALPHA-DUO-M1: 1200 mm²; ALPHA-MEGA-M1: 3000 mm² (consult factory)Built-in RCAP pressure controller for the ALPHA-DUO-M1 module(s) only (requires optional biased sample holders); not available for the ALPHA-MEGA-M1 module(s)1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-3M3Three ALPHA-MEGA-M1 modules2106 mm3000 mm² (consult factory)RCAP recoil-protection option is for ALPHA-DUO-M1 modules only; biased sample tray not available for Mega modules3Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-44Two ALPHA-DUO-M1 modules251 mm1200 mm²Built-in pressure controller; optional biased sample trays required1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-4M4Four ALPHA-MEGA-M1 modules2106 mm3000 mm² (consult factory)RCAP recoil-protection option is for ALPHA-DUO-M1 modules only; biased sample tray not available for Mega modules3Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-66Three ALPHA-DUO-M1 modules251 mm1200 mm²Built-in pressure controller; optional biased sample trays required1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-ENSEMBLE-88Four ALPHA-DUO-M1 modules251 mm1200 mm²Built-in pressure controller; optional biased sample trays required1Computer control per moduleUSB type BMAESTRO included; AlphaVision is separate software––Up to 4 Duo/Mega modules
ALPHA-MEGA1Benchtop unit with one large chamber for samples up to 106 mm1106 mm3000 mm² (consult factory)External RCAP-2 required for recoil protection1Computer controlledUSB type BMAESTRO included; AlphaVision is separate software8.5 kg––
ALPHA-MEGA-M11Single-input Alpha Ensemble expansion module (one large chamber, samples up to 106 mm)1106 mm3000 mm² (consult factory)RCAP recoil-protection option is for ALPHA-DUO-M1 modules only; biased sample tray not available for Mega modules3Computer control per module–––Expansion module for Alpha Ensemble; not a standalone benchtop instrument–

Documents

Sources

The number after each value refers to the source listed below. Values are as published by the manufacturer; the exact configuration is always confirmed in our quotation.

  1. 1
    Publisher: ORTEC (AMETEK)Brochure p. 2 model descriptions; p. 6 Ordering Information; p. 3, Specifications – Vacuum Chamber; p. 3, Specifications – System Performance; p. 3, Electronics – Bias Supply … Checked 2026-10-16
  2. 2
    Publisher: ORTEC (AMETEK)Product catalogue PDF p. 9 (printed p. 6), 'Alpha Suite Alpha Spectrometers' ordering table; PDF p. 9 (printed p. 6), 'Alpha Suite Alpha Spectrometers' ordering table, 'Expansion Modules for Alpha Ensemble' Checked 2026-10-16
  3. 3
    Publisher: ORTEC (AMETEK)Manual PDF p. 35 (printed p. 27), section 5.8 Vacuum Chamber Checked 2026-10-16
A-576 A-PAD alfaspektrometrikedja
ORTEC

A-576 A-PAD alfaspektrometrikedja

NIM alpha spectroscopy chain

Selectable energy ranges
3–8; 4–7; 3–5; 4–6; 5–7; 6–8 MeV
Net weight
0.9 kg
12 referenced values
Contact us

Questions about this product?

We help you choose the right configuration, detector and accessories for your measurement task.